Industrial Test Systems 486691 Скачать руководство пользователя страница 14

Iron (Fe

+2

/Fe

+3

) Table

 Iron results require the table below. Follow 

eXact

®

 Micro 7+ Iron (Fe

+2

/Fe

+3

) Test Procedure

 (above) using 

eXact

®

Reagent Iron Ferro Powder Pillow, Part No. 481623

eXact

®

 Reagent Iron Ferro Powder Pillow, Part No. 481623 - for 4mL Samples

Rev. 012809 TFE

tr

90
80
70
60
50
40
30
20
10

0

9

8

7

6

5

4

3

2

1

0

<0.15

0.25
0.55
0.87
1.24
1.69
2.26

3.1

>3.5
>3.5

<0.15

0.29
0.58
0.90
1.28
1.74
2.32

3.2

>3.5
>3.5

6

Find the “TR” result in the table below to determine the Iron concentration in ppm (parts per
million). (Example: a “TR” result of 85.3 (use only the 85 for the chart) equals an Iron value of
0.37 ppm). Record result. After testing is completed, rinse cell immediately.

USE TABLE

4

Press the 

ZERO/ON

 button. The cursor will move across the display, followed by 

100 %T

.

Sample is ready for testing.

ZERO METER

1

Press the 

ZERO/ON

 button to power the meter on; the display will show all annunciators,

then the current MENU selection, followed by the last reading.

TURN METER ON

2

Press and re-press the 

MENU

 button until the display shows the parameter TR7.

SELECT TEST: TR7

3

Rinse the 

CELL

 at least 3 times with the water sample you will be testing - rinsing minimizes

the potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml)
with the water sample. In order to leave room for powder reagent, do not overfill cell.

FILL METER WITH SAMPLE

Iron (Fe

+2

/Fe

+3

) Test Procedure

<0.15

0.32
0.61
0.94
1.32
1.79
2.39

3.4

>3.5
>3.5

<0.15

0.35
0.64
0.97
1.36
1.84
2.46

3.5

>3.5
>3.5

<0.15

0.37
0.67
1.01
1.41
1.89

2.5

>3.5
>3.5
>3.5

<0.15

0.40
0.70
1.05
1.45
1.95

2.6

>3.5
>3.5
>3.5

0.15
0.43
0.74
1.09
1.49
2.00

2.7

>3.5
>3.5
>3.5

0.18
0.46
0.77
1.12
1.54
2.07

2.8

>3.5
>3.5
>3.5

0.21
0.49
0.80
1.16
1.59
2.13

2.9

>3.5
>3.5
>3.5

0.24
0.52
0.84
1.20
1.64
2.19

3.0

>3.5
>3.5
>3.5

14

5

Add the contents of one 

eXact

®

 Reagent Iron Ferro Powder Pillow, Part No. 481623

 to the

CELL

 and cap meter cell with mixing cap. Press 

READ

 to start the timer, mix the sample by

pressing one finger over the cap to keep it in place. Then, turn the meter over from side to side
to mix during the 20 second countdown. 

When time displays 1

, hold the meter upright and the

cursor will move across the display, informing you that it is about to measure the sample. Time
the reaction in the cell for an additional 

three (3) minutes

 (timer not included). During this time,

meter will shut off. When 

3 minutes

 have elapsed, turn meter on and wait for the display to

show last reading. Then, press 

READ

, which will start a final 

20 SECOND

 countdown. The

cursor will move across the display, informing you that it is about to measure the sample.
Record result displayed (this result is automatically stored in TR7).

ADD REAGENT, CAP, PRESS “READ”, AND MIX

FE

TR7

NOTE:

 Water samples with suspended solids and particulates may cause incorrect results, since the reducing agent in the 

powder pillow often dissolves the suspended solids and particulates. If the meter is zeroed with the original water sample 
the zero is no longer valid for this kind of sample. So, for this kind of sample, zero the meter using a water sample free of 
suspended solids and particulates (use distilled, deionized, or bottled water in Step 4). Then, rinse the cell three times with 
the water sample you will be testing, and finally fill the cell with this water sample. Continue with Steps 5 and 6 as directed.

Fluoride Table

 Fluoride results require the table below. Follow 

eXact

®

 Micro 7+ Fluoride (SPADNS) Test

Procedure

 (above) using 

eXact

®

 Reagent F

 (SPADNS), Part No. 486643.

eXact Reagent F

 (SPADNS), Part No. 486643 - for 4mL Samples

Rev. 121208 F

tr

4
3
2
1

0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1  0

>1.1

1.1
0.7
0.2

>1.1

1.1
0.6
0.2

8

Find the “TR” result in the table below to determine the Fluoride concentration in ppm (parts per
million). (Example: a “TR” result of 3.42 (use only the 3.4 for the chart) equals a Fluoride value
of 0.9 ppm). Record result. After testing is completed, rinse cell immediately. For all values
with a Transmission above 5.00%, the Fluoride level is greater than 1.1 ppm. For all samples
above 1.1 ppm, it is recommended that the sample is diluted with distilled or deionized water
and retested.

USE TABLE

7

Press READ to start timer, mix sample by pressing one finger over the cap to keep it in place. Then,
gently turn the meter over from side to side to mix during the 

20 SECOND

 countdown. 

When time

displays 1

, hold meter upright and the cursor will move across the display, informing you that it is

about to measure the sample. Record result displayed (this result is automatically stored in TR7).

PRESS READ

6

Fill the cell to the top with the sample to be tested (do this slowly to prevent overfilling). Cap the
meter with the cell mixing cap.

ADD SAMPLE AND CAP

4

Press the 

ZERO/ON

 button. The cursor will move across the display, followed by 

100 %T

.

Discard the sample (empty cell).

ZERO METER

1

Press the 

ZERO/ON

 button to power the meter on; the display will show all annunciators, then

the current MENU selection, followed by the last reading.

TURN METER ON

2

Press and re-press the 

MENU

 button until the display shows the parameter TR7.

SELECT TEST: TR7

3

Rinse the 

CELL

 at least 3 times with the water sample you will be testing - rinsing minimizes

the potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml) with
the water sample.

FILL METER WITH SAMPLE

Fluoride 

(SPADNS)

 Test Procedure

>1.1

1.0
0.6
0.2

>1.1

1.0
0.5
0.1

>1.1

0.9
0.5
0.1

>1.1

0.9
0.4

<0.1

>1.1

0.8
0.4

<0.1

>1.1

0.8
0.4

<0.1

>1.1

0.7
0.3

<0.1

>1.1

0.7
0.3

<0.1

19

5

Shake the bottle of 

eXact

®

 Reagent F

 (SPADNS), Part No. 486643

 to mix chemical in the

bottle. Then, add ten (10) drops of eXact

®

 Reagent F

 (SPADNS) to the empty cell.

ADD REAGENT

F

TR7

Содержание 486691

Страница 1: ...3 329 9743 ITS SENSAFE COM WWW SENSAFE COM www poolcheckonline com For European Middle East Inquiries and Re Orders ITS Europe LTD Unit B Stanley Crt Glenmore Business Park Telford Rd Churchfields Industrial Estate Salisbury Wiltshire SP2 7GL United Kingdom Tel 44 0 1722 329502 Fax 44 0 1722 329880 SALES SENSAFE COM WWW ITSEUROPE CO UK www poolcheckonline com eXact Strip Micro DPD 1 Accuracy eXact...

Страница 2: ...7 to be run first Alizarin Red S Citrate DPD Oxalic Acid DPD Kl DPD Kl Biquinoline DPD Kl DPD Phenol Red Salicylate Method Silver ppt DPD Diphenylcarbazide Melamine ppt SPADNS DPD TPTZ 1 10 Phenanthroline Phthalein Purple PAN Cyanide Zinc Reduction Chromotropic Acid Alizarin Red S Thymol Blue Tetraphenylborate Barium ppt Turbidity values as NTU PARAMETER No PART NUMBER OF TESTS DETECTION RANGE CHE...

Страница 3: ...0 The following are some common messages that may be displayed including error messages If an error message other than those listed below is displayed please contact technical support in the USA at 803 329 0162 ext 0 LCD Message HI LO LO ER Description In READ mode test sample concentration is above the measurement range test specific In READ mode test sample concentration is below the measurement...

Страница 4: ... for pH and Total Alkalinity Sample water that may splash out of the CELL during movement of the eXact Strip Micro will not affect accuracy as long as the cell is 50 full at end of test Rinse the CELL with clean water immediately after completing the test Some test reagents will stain or coat the CELL Just before testing rinse the sample CELL with the sample water several times to get a representa...

Страница 5: ... time the KI is added For compliance testing you must time the two minutes and then make your measurement NOTE From testing in our lab water samples above 70 F 20 C generally reach a stabilized reading quickly eXact Strip Micro CL DPD 1 DPD 3 DPD 4 Interferences part nos 486637 486638 486670 Interfering Substance Acidity Alkalinity Bromine Bromamines Br2 Chlorine Dioxide ClO2 Copper Cu 2 Iodine I2...

Страница 6: ...y 0 01 press READ again and repeat After testing is completed rinse cell immediately Record the Total Chlorine as the highest value the meter displayed NOTE Standard Method 4500 Cl G procedure for total chlorine requires the reading to be made 2 minutes after the reagents are added pH Test Procedure 1 Remove one 1 eXact Strip Micro PH part number 486639 from the bottle before beginning the test Se...

Страница 7: ...n the display will show all annunciators then the current MENU selection followed by the last reading TURN METER ON 3 Press and re press the MENU button until the display shows the parameter AL4 SELECT TEST AL4 4 Rinse the CELL at least 3 times with the water sample you will be testing rinsing minimizes the potential for cross contamination from a previous test Finally fill cell to capacity 4ml wi...

Страница 8: ...TER WITH SAMPLE Copper Test Procedure CA5 Calcium reacts with Oxalic acid as it is released from the strip to form a white precipitate directly proportional to the Calcium concentration Values are reported as Calcium Carbonate CU6 Copper reacts with Biquinoline or Bicinchoninic Acid as it is released from the strip to form a purple color directly proportional to the copper concentration CA CA5 CU ...

Страница 9: ... charts round off your T value to two digits Tests that can be used with the Micro 7 in TR7 MENU are listed on page 31 Typically the most common tests are supplied with the conversion chart in this booklet For conversion charts and information about the tests not given in this booklet visit our website You can also use the TR7 MENU to develop your own custom conversion chart for unusual samples to...

Страница 10: ...TER WITH DISTILLED OR DEIONIZED WATER Turbidity Test Procedure TU TR7 Alkali pH Table Alkali pH results require the table below Follow eXact Micro 7 Alkali pH Test Procedure above using eXact Strip Micro Alkali pH Part No 486609 eXact Strip Micro Alkali pH Part No 486609 for 4mL Samples Rev 020509 AlkpH tr 90 80 70 60 50 40 30 20 10 0 9 8 7 6 5 4 3 2 1 0 8 3 8 3 8 3 8 3 8 5 8 7 8 9 9 1 9 3 9 6 8 3...

Страница 11: ...14 18 23 29 39 60 5 5 8 11 15 19 24 30 40 60 5 5 8 12 15 19 24 31 42 60 5 6 9 12 15 20 25 31 45 60 Cyanuric Acid Test Procedure CY TR7 Iron as Fe 2 Table Iron results require the table below Follow eXact Micro 7 Iron as Fe 2 Test Procedure above using eXact Strip Micro FE Part No 486631 6 Find the TR result in the table below to determine the Iron concentration in ppm parts per million Example a T...

Страница 12: ...0 03 0 03 0 39 1 06 1 84 2 8 4 1 5 8 5 8 5 8 12 NH3 TR7 Chlorine Dioxide Table Chlorine Dioxide results require the table below Follow eXact Micro 7 Chlorine Dioxide Test Procedure above using eXact Strip Micro CL DPD 1 Part No 486637 eXact Strip Micro CL DPD 1 Part No 486637 for 4mL Samples Rev 121908 ClO2 tr 90 80 70 60 50 40 30 20 10 0 9 8 7 6 5 4 3 2 1 0 0 03 0 23 0 51 0 82 1 19 1 62 2 2 2 9 4...

Страница 13: ...hloride as NaCl Test Procedure For Tap Water CHLORIDE TR7 6 Dip the eXact Strip Micro CHLORIDE as NaCl Part No 481657 into the CELL and immediately press READ This starts the 20 SECOND countdown timer During this time move the strip in a gentle back and forth motion Remove and discard the strip when the time displays 1 The cursor will move across the display informing you that it is about to measu...

Страница 14: ...ples with suspended solids and particulates may cause incorrect results since the reducing agent in the powder pillow often dissolves the suspended solids and particulates If the meter is zeroed with the original water sample the zero is no longer valid for this kind of sample So for this kind of sample zero the meter using a water sample free of suspended solids and particulates use distilled dei...

Страница 15: ... 7 Find the TR result in the table below to determine the Manganese concentration in ppm parts per million Example a TR result of 65 3 use only the 65 for the chart equals a Manganese value of 0 04 ppm Record result After testing is completed rinse cell immediately USE TABLE 4 Press the ZERO ON button The cursor will move across the display followed by 100 T Sample is ready for testing ZERO METER ...

Страница 16: ...8 39 62 80 100 121 140 140 140 140 8 41 64 82 102 124 140 140 140 140 8 43 66 84 104 126 140 140 140 140 13 46 68 86 106 128 140 140 140 140 18 49 70 88 108 131 140 140 140 140 NOTE Divide the above Nitrate result by 4 4 to determine Nitrate value as Nitrogen NO3 as N 16 NO3 TR7 Nitrite NO2 Table Nitrite results require the table below Follow eXact Micro 7 Nitrite NO2 Test Procedure above using eX...

Отзывы: