Industrial Test Systems 486691 Скачать руководство пользователя страница 1

Index:

Parameter Page 

MENU

Free Chlorine (DPD-1) 

CL1

Permanganate (DPD-1) 

CL1

Total Chlorine (DPD-3) 

CL1

Ozone (DPD-4) 

CL1

Total Chlorine (DPD-4) 

CL1

pH 6 

PH2

Bromine 7 

BR3

Total Alkalinity 

AL4

Calcium Hardness (as CaCO

3

) 8 

CA5

Copper 8 

CU6

(NOTE: The below tests are read as %Transmission and
require use of a conversion table)

General TR7 Procedure 

Tr7

Turbidity 10 

Tr7

Cyanuric Acid 

11 

Tr7

Ammonia 12 

Tr7

Chloride (in Tap Water) 

13 

Tr7

Iron (as Fe

+2

 / Fe

+3

) 14 

Tr7

LR Total Hardness (up to 77ppm) 15 

Tr7

Nitrate 16 

Tr7

Nitrite 17 

Tr7

Manganese 18 

Tr7

Fluoride 19 

Tr7

Chloride (Pool Salt Systems) 

20 

Tr7

Chlorine Dioxide (DPD-1) 

21 

Tr7

Iron (as Fe

+2

) 22 

Tr7

Alkali pH 

23 

Tr7

Acid pH 

24 

Tr7

Chromate (Chromium) 

25 

Tr7

Sulfate (SO

4

) 26 

Tr7

Specifications 2
About Your Photometer 

3

Notes 27-28
Tips for Best Accuracy 

29

Battery Installation 

30

Warranty (2 year) 

30

Reorder Information 

31

USEPA Compliance 

2, 32

Kit Components 

32

Visit us online at 

sensafe.com/

 

for up-to-date

product information & NEW tests available.

IDEAL FOR DRINKING WATER, POOLS AND SPAS,

ENVIRONMENTAL, & EDUCATIONAL TESTING

Revision 03/05/09

486691-K

 Standard Kit

486691-KP

 Pool/Spa Kit

486691-WD

 Well Driller Kit

The eXact

®

 Micro 7+ Advanced 

Photometer System has been 

designed for use with the eXact

®

 Strip 

Micro reagent delivery system.

© 2009 Industrial Test Systems, Inc.  •  EXACT

®

 is a registered trademark of Industrial Test Systems, Inc., Rock Hill, SC USA

Micro 7+

Advanced Photometer System

U.S. Patent No. 7,333,194, U.S. Patent No.  7,491,546, South African Patent No. 2007/0628 and international

patent applications including International Patent Appln. No. PCT/US2005/033985; and Eur. Pat. App. 1,725,864

USEPA, DIN, & ISO Compliant for Free & Total Chlorine Testing

(4500-CL G, DIN Standard 38 408 G4, ISO 7393/2)

Manufactured By:

 Industrial Test Systems, Inc.

1875 Langston Street, Rock Hill, SC 29730 USA

Phone:

 1-800-861-9712 

- INSIDE THE U.S.

1-803-329-9712 

- OUTSIDE THE U.S.

Fax:

 1-803-329-9743

[email protected]

WWW.SENSAFE.COM

www.poolcheckonline.com

P

F

Micro 7+ Ad

m ha

19

uding Inter

32

For US Inquiries and Re-Orders:

Industrial Test Systems, Inc.

1875 Langston Street,
Rock Hill, SC  29730 USA

Phone:

 1-800-861-9712 - 

INSIDE THE U.S.

1-803-329-9712 - 

OUTSIDE THE U.S.

Fax:

 1-803-329-9743

[email protected]

WWW.SENSAFE.COM

www.poolcheckonline.com

For European & Middle East Inquiries and Re-Orders:

ITS Europe, LTD

Unit B Stanley Crt., Glenmore Business Park, Telford Rd.
Churchfields Industrial Estate,
Salisbury, Wiltshire SP2 7GL, United Kingdom

Tel:

 +44 (0)1722 329502 

Fax:

 +44 (0) 1722 329880

[email protected]

WWW.ITSEUROPE.CO.UK

www.poolcheckonline.com

eXact

®

 Strip Micro DPD-1 Accuracy

eXact Strip DPD-1 on Hach DR890 and eXact Micro 7

+

y = 0.9703x + 0.0561

R

2

 = 0.9912

0.00

1.00

2.00

3.00

4.00

5.00

6.00

0.00

1.00

2.00

3.00

4.00

5.00

6.00

Hach DR890 (ppm)

eXact

 Micro

 7+

ppm (mg/L)

ppm (mg/L)

DR890

Micro 7+

0.00

0.01

0.04

0.03

0.07

0.08

0.79

0.73

0.98

0.96

1.55

1.52

2.08

2.12

2.4

2.6

2.9

3.3

4.0

3.9

5.5

5.2

Free Chlorine results are compared using the 

eXact

®

 Strip Micro CL (DPD-1)

 with the eXact

®

 Micro 7+ Meter in

Menu CL1 and Hach

®

 DR890 Colorimeter in Program 9 and Program 12 using Hach

®

 powder pillows.

Meter

Menu

Range (PPM)

Resolution

Micro 7+ CL1

0.00 to 2.39

0.01

2.4 to 6.0

0.1

DR890

Program 9

0.00 to 2.20

0.01

Program 12

0.0 to 11.0

0.1

Hach

®

 is a registered trademark of Danaher Corporation

(486691-K)

Standard Kit 

Includes:

1 eXact

®

 Micro 7+ Meter 

(486691)

eXact

®

 Strip Micro DPD-1 

(486637-25)

eXact

®

 Strip Micro DPD-3 

(486638-25)

eXact

®

 Strip Micro pH 

(486639-25)

eXact

®

 Strip Micro Total Alkalinity 

(486641-25)

eXact

®

 Strip Micro Calcium Hardness 

(486629-25)

eXact

®

 Strip Micro Copper 

(486632-25)

1 Mixing Cap
1 Cell Cleaning Brush
This Instruction Booklet
Plastic Carrying Case
AAA Batteries (4)

(486691-KP)

Pool/Spa Kit

Includes:

1 eXact

®

 Micro 7+ Meter 

(486691)

eXact

®

 Strip Micro DPD-1 

(486637-25)

eXact

®

 Strip Micro DPD-3 

(486638-25)

eXact

®

 Strip Micro pH 

(486639-25)

eXact

®

 Strip Micro Total Alkalinity 

(486641-25)

eXact

®

 Strip Micro Calcium Hardness 

(486629-25)

eXact

®

 Strip Micro Copper 

(486632-25)

eXact

®

 Reagent Cyanuric Acid 

(481652)

1 Mixing Cap
1 Cell Cleaning Brush
This Instruction Booklet
Plastic Carrying Case
AAA Batteries (4)

(486691-WD)

Well Driller Kit

Includes:

1 eXact

®

 Micro 7+ Meter 

(486691)

eXact

®

 Strip Micro DPD-1 

(486637-25)

eXact

®

 Strip Micro DPD-3 

(486638-25)

eXact

®

 Strip Micro pH 

(486639-25)

eXact

®

 Strip Micro Total Alkalinity 

(486641-25)

eXact

®

 Strip Micro Calcium Hardness 

(486629-25)

eXact

®

 Strip Micro Copper 

(486632-25)

eXact

®

 Reagent Iron 

(481623)

eXact

®

 Micro Manganese 

(486606)

1 Mixing Cap
1 Cell Cleaning Brush
This Instruction Booklet
Plastic Carrying Case
AAA Batteries (4)

The eXact

®

 Micro 7+ Line of Kits

Contact Information

Содержание 486691

Страница 1: ...3 329 9743 ITS SENSAFE COM WWW SENSAFE COM www poolcheckonline com For European Middle East Inquiries and Re Orders ITS Europe LTD Unit B Stanley Crt Glenmore Business Park Telford Rd Churchfields Industrial Estate Salisbury Wiltshire SP2 7GL United Kingdom Tel 44 0 1722 329502 Fax 44 0 1722 329880 SALES SENSAFE COM WWW ITSEUROPE CO UK www poolcheckonline com eXact Strip Micro DPD 1 Accuracy eXact...

Страница 2: ...7 to be run first Alizarin Red S Citrate DPD Oxalic Acid DPD Kl DPD Kl Biquinoline DPD Kl DPD Phenol Red Salicylate Method Silver ppt DPD Diphenylcarbazide Melamine ppt SPADNS DPD TPTZ 1 10 Phenanthroline Phthalein Purple PAN Cyanide Zinc Reduction Chromotropic Acid Alizarin Red S Thymol Blue Tetraphenylborate Barium ppt Turbidity values as NTU PARAMETER No PART NUMBER OF TESTS DETECTION RANGE CHE...

Страница 3: ...0 The following are some common messages that may be displayed including error messages If an error message other than those listed below is displayed please contact technical support in the USA at 803 329 0162 ext 0 LCD Message HI LO LO ER Description In READ mode test sample concentration is above the measurement range test specific In READ mode test sample concentration is below the measurement...

Страница 4: ... for pH and Total Alkalinity Sample water that may splash out of the CELL during movement of the eXact Strip Micro will not affect accuracy as long as the cell is 50 full at end of test Rinse the CELL with clean water immediately after completing the test Some test reagents will stain or coat the CELL Just before testing rinse the sample CELL with the sample water several times to get a representa...

Страница 5: ... time the KI is added For compliance testing you must time the two minutes and then make your measurement NOTE From testing in our lab water samples above 70 F 20 C generally reach a stabilized reading quickly eXact Strip Micro CL DPD 1 DPD 3 DPD 4 Interferences part nos 486637 486638 486670 Interfering Substance Acidity Alkalinity Bromine Bromamines Br2 Chlorine Dioxide ClO2 Copper Cu 2 Iodine I2...

Страница 6: ...y 0 01 press READ again and repeat After testing is completed rinse cell immediately Record the Total Chlorine as the highest value the meter displayed NOTE Standard Method 4500 Cl G procedure for total chlorine requires the reading to be made 2 minutes after the reagents are added pH Test Procedure 1 Remove one 1 eXact Strip Micro PH part number 486639 from the bottle before beginning the test Se...

Страница 7: ...n the display will show all annunciators then the current MENU selection followed by the last reading TURN METER ON 3 Press and re press the MENU button until the display shows the parameter AL4 SELECT TEST AL4 4 Rinse the CELL at least 3 times with the water sample you will be testing rinsing minimizes the potential for cross contamination from a previous test Finally fill cell to capacity 4ml wi...

Страница 8: ...TER WITH SAMPLE Copper Test Procedure CA5 Calcium reacts with Oxalic acid as it is released from the strip to form a white precipitate directly proportional to the Calcium concentration Values are reported as Calcium Carbonate CU6 Copper reacts with Biquinoline or Bicinchoninic Acid as it is released from the strip to form a purple color directly proportional to the copper concentration CA CA5 CU ...

Страница 9: ... charts round off your T value to two digits Tests that can be used with the Micro 7 in TR7 MENU are listed on page 31 Typically the most common tests are supplied with the conversion chart in this booklet For conversion charts and information about the tests not given in this booklet visit our website You can also use the TR7 MENU to develop your own custom conversion chart for unusual samples to...

Страница 10: ...TER WITH DISTILLED OR DEIONIZED WATER Turbidity Test Procedure TU TR7 Alkali pH Table Alkali pH results require the table below Follow eXact Micro 7 Alkali pH Test Procedure above using eXact Strip Micro Alkali pH Part No 486609 eXact Strip Micro Alkali pH Part No 486609 for 4mL Samples Rev 020509 AlkpH tr 90 80 70 60 50 40 30 20 10 0 9 8 7 6 5 4 3 2 1 0 8 3 8 3 8 3 8 3 8 5 8 7 8 9 9 1 9 3 9 6 8 3...

Страница 11: ...14 18 23 29 39 60 5 5 8 11 15 19 24 30 40 60 5 5 8 12 15 19 24 31 42 60 5 6 9 12 15 20 25 31 45 60 Cyanuric Acid Test Procedure CY TR7 Iron as Fe 2 Table Iron results require the table below Follow eXact Micro 7 Iron as Fe 2 Test Procedure above using eXact Strip Micro FE Part No 486631 6 Find the TR result in the table below to determine the Iron concentration in ppm parts per million Example a T...

Страница 12: ...0 03 0 03 0 39 1 06 1 84 2 8 4 1 5 8 5 8 5 8 12 NH3 TR7 Chlorine Dioxide Table Chlorine Dioxide results require the table below Follow eXact Micro 7 Chlorine Dioxide Test Procedure above using eXact Strip Micro CL DPD 1 Part No 486637 eXact Strip Micro CL DPD 1 Part No 486637 for 4mL Samples Rev 121908 ClO2 tr 90 80 70 60 50 40 30 20 10 0 9 8 7 6 5 4 3 2 1 0 0 03 0 23 0 51 0 82 1 19 1 62 2 2 2 9 4...

Страница 13: ...hloride as NaCl Test Procedure For Tap Water CHLORIDE TR7 6 Dip the eXact Strip Micro CHLORIDE as NaCl Part No 481657 into the CELL and immediately press READ This starts the 20 SECOND countdown timer During this time move the strip in a gentle back and forth motion Remove and discard the strip when the time displays 1 The cursor will move across the display informing you that it is about to measu...

Страница 14: ...ples with suspended solids and particulates may cause incorrect results since the reducing agent in the powder pillow often dissolves the suspended solids and particulates If the meter is zeroed with the original water sample the zero is no longer valid for this kind of sample So for this kind of sample zero the meter using a water sample free of suspended solids and particulates use distilled dei...

Страница 15: ... 7 Find the TR result in the table below to determine the Manganese concentration in ppm parts per million Example a TR result of 65 3 use only the 65 for the chart equals a Manganese value of 0 04 ppm Record result After testing is completed rinse cell immediately USE TABLE 4 Press the ZERO ON button The cursor will move across the display followed by 100 T Sample is ready for testing ZERO METER ...

Страница 16: ...8 39 62 80 100 121 140 140 140 140 8 41 64 82 102 124 140 140 140 140 8 43 66 84 104 126 140 140 140 140 13 46 68 86 106 128 140 140 140 140 18 49 70 88 108 131 140 140 140 140 NOTE Divide the above Nitrate result by 4 4 to determine Nitrate value as Nitrogen NO3 as N 16 NO3 TR7 Nitrite NO2 Table Nitrite results require the table below Follow eXact Micro 7 Nitrite NO2 Test Procedure above using eX...

Отзывы: