background image

Guidelines for using a motorized pump and sampling line 

When sampling with a motorized pump and sampling line, Industrial Scientific recommends the following: 

 

Choose the tubing type based on the target gases. If the target gases are known, use Teflon-lined 

tubing when sampling for these gases: chlorine (Cl

2

), chlorine dioxide (ClO

2

), hydrogen chloride (HCl), 

and volatile organic compounds (VOCs). For other known target gases, urethane tubing or Teflon-

lined tubing may be used.  

When the target gases are unknown, use Teflon-lined tubing.   

 

Know the length of the sample line as it is a factor in determining sampling time. A sample line may 

consist of tubing, a probe, or a probe and tubing. It should also have a dust filter–water stop installed 

at the line's end that will extend into the sample area. Sample-line length is defined as the distance 

from the dust filter–water stop opening to the point where the line connects to the pump's inlet. Ensure 

sample-line length does not exceed the pump's maximum draw.  

 

Before and after each air sample, perform a test of the full sampling line.  

o

 

Use a thumb to block the end of the sampling line at the water-stop opening. This should 

cause a pump-fault alarm.  

o

 

Remove the thumb from the water-stop opening. After the alarm cycle completes, the pump 

should resume normal operation.  

Note: If a pump fault does not occur, check and correct for cracks or other damage, debris, and proper installation in 

these areas: the sampling line and its connections, the pump's inlet cap and inlet barrel, and the dust filter-water stop 

items at the end of the sampling line and inside the pump inlet barrel.  

 

Based on sample-line length, calculate the minimum time recommended for the air sample to reach 

the instrument's sensors. As shown in Table 3, use a base time of 2 minutes, and add 2 seconds for 

each 30 cm (1 ') of line length. Watch the display screen for gas readings and, if present, allow them 

to stabilize to determine the reading.  

Table 1.3 Minimum sample time for common sample-line lengths 

Sample-line 

length 

Base time 

(minutes) 

Sample-line-length 

factor 

Minimum sample time 

(mm:ss) 

3.05 m (10 ') 

2 min 

(10 ' x 2 s) 

02:20 

6.10 m (20 ') 

2 min 

+

 

(20 ' x 2 s) 

02:40 

9.14 m (30 ') 

2 min 

+

 

(30 ' x 2 s) 

03:00 

12.10 m (40 ') 

2 min 

+

 

(40 ' x 2 s) 

03:20 

15.24 m (50 ') 

2 min 

(50 ' x 2 s) 

03:40 

18.29 m (60 ') 

2 min 

(60 ' x 2 s) 

04:00 

21.34 m (70 ') 

2 min 

(70 ' x 2 s) 

04:20 

24.38 m (80 ') 

2 min 

(80 ' x 2 s) 

04:40 

27.43 m (90 ') 

2 min 

(90 ' x 2 s) 

05:00 

30.48 m (100 ') 

2 min 

(100 ' x 2 s) 

05:20 

Содержание Ventis 17154524-1

Страница 1: ...Field Guide Setup Operation and Service Edition 8 May 14 2018 Part Number 17154524 1 ...

Страница 2: ...rial Scientific Corporation Pittsburgh PA USA Industrial Scientific Co Ltd Shanghai China 2016 2017 2018 Industrial Scientific Corporation All rights reserved Published 2018 Revision 10 www indsci com ventis vsp ...

Страница 3: ...trument installation and removal 8 Pump test and operation 9 Shutting down 9 Service 10 Door Installation and removal 10 Battery 10 Water barrier 10 Exhaust filter or audible alarm filter 10 Alarms 11 Warranty 12 Limitation of Liability 12 Contact Information 13 Tables and Figures Table 1 1 Certifications 1 Table 1 2 Warnings and cautionary statements 2 Table 1 3 Minimum sample time for common sam...

Страница 4: ......

Страница 5: ...NMETRO Ex ia IIC T4 Ga 20 C to 50 C 4 F to 122 F UL Class I Division 1 Groups A B C D T4 Class I Zone 0 AEx ia IIC T4 Ga Class II Group F G Carbonaceous and Grain Dust UL 913 8th Ed UL 60079 0 6th Ed UL 60079 11 6th Ed CSA C22 2 No 157 20 C to 50 C 4 F to 122 F China Ex Ex ia IIC T4 Ga GB 3836 1 2010 GB 3836 4 2010 GB 3836 20 2010 20 C to 50 C 4 F to 122 F The EC type examination certificate is DE...

Страница 6: ...e or charge the battery packs only in an area known to be nonhazardous Not for use in oxygen enriched atmospheres Battery contacts are exposed on battery packs when they are removed from the instrument Do not touch the battery contacts and do not stack battery packs on top of each other Insert the alkaline batteries with the correct positive and negative orientation The Ventis Slide on Pump is app...

Страница 7: ... the sampling line at the water stop opening This should cause a pump fault alarm o Remove the thumb from the water stop opening After the alarm cycle completes the pump should resume normal operation Note If a pump fault does not occur check and correct for cracks or other damage debris and proper installation in these areas the sampling line and its connections the pump s inlet cap and inlet bar...

Страница 8: ... hinged door is opened and secured closed by the body mounted latch therefore no tools are needed for instrument installation or removal The pump also allows for the installation of a clip equipped instrument Power source The pump is powered by its own battery and it does not draw power from the instrument battery When using Li ion battery packs the pump and instrument batteries should be charged ...

Страница 9: ... XX Air sample tubing kit Urethane Batteries 17134453 XY Li ion battery kit Run time is 18 hours at 20º C recharge time is 3 to 5 hoursª 17148313 Y Extended range Li ion battery pack Run time is 36 hours at 20º C recharge time is 3 to 7 5 hoursª 17151184 XY Cover For use with Extended Range Lithium ion battery pack 17154577 XY Alkaline battery kit Includes pack batteries and cover Run time 10 hour...

Страница 10: ......

Страница 11: ...ng or appears to have been damaged contact a local distributor of ISC products or ISC See Contact Information indicates part is not field replaceable See Contact Information to locate an ISC Service Center Setup Operation and Service Pump setup WARNING This task must be completed in an area known to be nonhazardous 1 Attach the battery to the upper portion of the pump case back the battery contact...

Страница 12: ...ump inlet door openings exhaust filter and audible alarm filter see inset photo of any dirt or debris 3 Toinstall the instrument place it in the pump cradle at a slight angle the instrument top meets the cradle top 4 Press down on the instrument to secure it in the cradle NOTE Remove the instrument in a similar fashion by pulling it up from and out of the pump Tip To prevent damage to or loss of t...

Страница 13: ...e sample tubing to the pump inlet s nipple Use a compatible water stop at the other end of the sampling line Operation at room temperature 4 Use a thumb to block the end of the sampling line at the water stop opening This should cause a pump fault alarm Remove the thumb from the water stop opening After the alarm cycle completes the pump should resume normal operation Note If a pump fault does not...

Страница 14: ... back align the battery contacts with the top of the pump Tighten the battery s four captive screws to a torque setting of 39 newton meters 55 ounce inch Water barrier 1 Power off the pump 2 Turn the pump inlet cap counterclockwise to remove 3 Remove the water barrier from the inlet barrel 4 Place the new water barrier inside the inlet barrel the side with the larger filter surface should face the...

Страница 15: ... barrier for blockage See an on site supervisor or contact ISC TechnicalSupport see Contact Information Pump flow fault Audible alarm turns on Each LED blinks once Audible alarm turns off The above sequence repeats continuously The pump nipple inlet barrel water barrier exhaust inside pump cradle or sample tubing is blocked Check for debris at the pump nipple inside the inlet barrel at the water b...

Страница 16: ...D OF THE POTENTIAL FOR SUCH DAMAGES Industrial Scientific s total liability hereunder from any cause whatsoever except liability from personal injury caused by Industrial Scientific s negligence whether arising under contract warranty tort including negligence strict liability products liability or any other theory of liability will be limited to the lesser of Buyer s actual damages or the price p...

Страница 17: ... 86 400 820 2515 To locate a nearby distributor of our products or an Industrial Scientific service center or business office visit us at www indsci com Rendez vous sur notre site Web www indsci com si vous voulez trouver un distributeur de nos produits près de chez vous ou si vous recherchez un centre de service ou un bureau Industrial Scientific Besuchen Sie uns unter www indsci com um einen Ver...

Отзывы: