background image

22

Shorting Wire

NOTE: The TMX412 must be calibrated whenever a

sensor is installed or replaced. All installed
sensor types are shown on the Sensor
Configuration screen during instrument start-up.
However, a fault condition will be indicated if the
instrument has not been calibrated for any of the
installed sensor types. When a new oxygen sensor
is installed, allow 15 minutes for it to stabilize
before attempting calibration. If the toxic sensor
is replaced with a different type or the
combustible sensor is reconfigured, the alarm
settings must be changed to reflect the
appropriate alarm settings for the sensor
installed.

8.5.1   TOXIC SENSOR REMOVAL/

REPLACEMENT

• To remove the toxic sensor, grasp the sensor and lift it

straight up. Use care to avoid bending the sensor pins.

WARNING: When removing the toxic sensor for

storage, connect a shorting wire to
the two pins as shown.

• New sensors are shipped with a shorting wire attached.

Remove the shorting wire from the new sensor.

• Immediately install the sensor in the instrument. Never

apply pressure to the area inside of the black O-Ring seal
at the top of the sensor.

8.5.2   COMBUSTIBLE SENSOR REMOVAL/

REPLACEMENT

• To remove the combustible sensor, grasp the sensor and

lift it straight up. 

• When installing the combustible sensor, align the notch

on the sensor body with the polarizing pin on the 
sensor board.

8.5.3   OXYGEN SENSOR REMOVAL/

REPLACEMENT

• Remove the two mounting screws and nylon washers

using the maintenance tool or small Phillips screwdriver.

• Bend the two mounting tabs out as shown before

installing the new sensor.

!

23

OXYGEN

LEL

Toxic Sensor 1

Toxic Sensor 2

Polarizing Pin

Combustible Gas
Sensor

Oxygen Sensor

Easy-to read
Liquid Crystal
Display

8.6   I

NSTRUMENT

R

EASSEMBLY

A

FTER

S

ERVICE

• Reassemble the instrument by aligning the two posts

with the corresponding openings in the circuit board.
Gently fit the instrument case halves together, making
certain that the gasket is properly seated, and the flex
cable is not pinched between the buzzer and the
instrument case top.

• Replace and tighten the three retaining screws until they

are firmly seated. Do not overtighten.

• Replace the battery and battery cover. 

Содержание TMX412

Страница 1: ......

Страница 2: ...rinted 1000 5000 Specifications Subject to Change 1001 Oakdale Road Oakdale PA 15071 1500 412 788 4353 Toll Free 1 800 DETECTS FAX 412 788 8353 Multi Gas Monitor TMX412 Instruction Manual G U A R A N...

Страница 3: ...OUR MISSION Design Manufacture Sell Highest quality products for the preservation of life and property Provide Best customer service available...

Страница 4: ...uction Manual it will provide you with years of reliable monitoring I am most concerned that you be pleased with the performance of your TMX412 in the months and years ahead I urge you to call us with...

Страница 5: ...en mixtures only CAUTION High Over Range OR combustible gas readings may indicate an explosive concentration of combustible gas Multi Gas Monitor TMX412 1 WARNINGS AND CAUTIONARY STATEMENTS 3 2 UNPACK...

Страница 6: ...sociation has assessed only the combustible gas portion of this instrument for performance Mine Safety and Health Administration MSHA Workcover Authority NSW Australia Department of Mineral Resource N...

Страница 7: ...ygen Sensor Display Battery Condition Display E Backlight Key Toxic Sensor Port 2 Combustible Gas Sensor Port Combustible Sensor Display Toxic Sensor Display Unmarked keys allow alarm and calibration...

Страница 8: ...splay the type of sensor in the position in which it is installed in the instrument HOLD RELEASE 8 4 INSTRUMENT OPERATION 4 1 CHARGING THE BATTERY CAUTION If the TMX412 is using the nickel cadmium bat...

Страница 9: ...m the combustible gas sensor to prevent damage due to the high level of combustible gas O 0 21 0 OR 10 4 4 3 ZERO This mode allows the user to zero the instrument and calibrate all installed sensors R...

Страница 10: ...nstrument in clean air is preferred provided that there is no trace of toxic or combustible gas that would cause the instrument to respond If the air purity is uncertain use a cylinder of zero grade a...

Страница 11: ...to CAL will scroll across the bottom of the display Press E to continue calibrating the remaining sensors Ignoring this message will result in the instrument returning to the normal operating mode in...

Страница 12: ...ll sensors The display will show the first span value eg 25 LEL along with the prompt PRESS FOR NEXT E TO SELECT Press E to select the value you wish to change The display will flash the current span...

Страница 13: ...HTED AVERAGE The accumulated gas exposure averaged over a predetermined time typically eight hours 7 1 8 STEL SHORT TERM EXPOSURE LIMIT The accumulated gas exposure value averaged for the proceeding f...

Страница 14: ...When changing lithium battery cells replace the cells within the lithium battery pack in a non hazardous location Replace the lithium cells with three Duracel DL123A or three Panasonic CR123A or thre...

Страница 15: ...tely install the sensor in the instrument Never apply pressure to the area inside of the black O Ring seal at the top of the sensor 8 5 2 COMBUSTIBLE SENSOR REMOVAL REPLACEMENT To remove the combustib...

Страница 16: ...24 25...

Страница 17: ...x 25 SST 2 34 1704 2219 Screw FHCR 4 40 x 38 SST 3 35 1704 9736 Conductive Tape RF Screen 1 36 1705 0776 Circuit Board Support 2 37 1704 3308 Battery Spacer 1 38 1704 4066 Data Port Plug 1 39 1704 676...

Страница 18: ...ompact Charger Australian Version 1810 2231 230 VAC Two Unit Compact Charger 1810 1972 230 VAC One Unit Dual Rate Charger 1810 2145 12 VDC One Unit Compact Charger 1810 1915 12 VDC Two Unit Dual Rate...

Страница 19: ...RIGINAL PURCHASE PRICE OF THE NON CONFORMING GOODS IN NO EVENT WILL INDUSTRIAL SCIENTIFIC BE LIABLE FOR ANY OTHER SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES INCLUDING LOSS OF PROFIT OR LOSS OF USE AR...

Отзывы: