Industrial Scientific TMX412 Скачать руководство пользователя страница 11

15

SETSPAN

PRESS

INDUSTRIAL 

SCIENTIFIC

TMX412

OXYGEN

LEL/CH4

E

BATT

0

2 10

0

0

*

6.

C

HANGING

I

NSTRUMENT

S

ETTINGS

Instrument settings, including alarm values, calibration gas
concentrations and security code, may only be accessed
and changed during the instrument startup sequence.

A keypad overlay is supplied to reveal the location of the
hidden keys used for changing instrument settings.
Remove the overlay from the protective backing and place
it over the instrument faceplate. The clear portion of the
overlay will cling to the display window of the faceplate
and may be reused.

To access the instrument menus:

• Turn the TMX412 off and back on again.
• When the display shows the warm up timer, press the

plus (+) and minus (-) keys simultaneously.

If the instrument security code has been set to a value
other than “0”, the CODE screen will be displayed along
with the scrolling prompt PRESS (+) OR (-) TO SET (E)
TO ENTER. Use the (+) and (-) keys to input the correct
security code value and press (E). When the correct code
has been entered successfully, the instrument will
immediately enter the settings mode.

The settings mode consists of three main functions in the
following order:

SETSPAN
ALARMS
BATTERY
CODE

The scrolling prompt, PRESS (+) FOR NEXT (E) TO
SELECT appears on each screen. Press the (+) to step
through the list and (E) to select any of the main functions.
Pressing the MODE switch at any one of the main
function headings will cause the instrument to return to the
normal operating mode.

6.1   S

ETSPAN

The SETSPAN function allows the user to set the
combustible sensor monitoring range to either LEL or CH

4

and to set the calibration gas values for the combustible
and toxic sensors.

• Press (E) to enter the SETSPAN function. The display

will show LEL along with the prompt PRESS (+) TO
CHANGE.

14

• At the completion of the oxygen sensor span operation

the instrument will emit a short beep and the display will
indicate GO CAL and the message PRESS (E) to CAL
will scroll across the bottom of the display. Press (E) to
continue calibrating the remaining sensors. Ignoring this
message will result in the instrument returning to the
normal operating mode in approximately eight seconds.

• When the GO CAL prompt has been acknowledged, the

instrument will display the first sensor to be calibrated
along with the span gas value. The message APPLY CAL
GAS will scroll across the bottom of the display.

NOTE: APPLY CAL GAS means connect the calibration

cup to the appropriate calibration gas cylinder.
Place the calibration cup on the instrument as
shown. Turn on gas supply.

• Apply the appropriate concentration of calibration gas.

The instrument will wait for three minutes to sense that
calibration gas has been applied before aborting and
failing calibration. When the instrument senses a
concentration greater than 50% of the span gas setting
the display will indicate the current full span value for
that sensor. The message CAL IN PROCESS will scroll
across the bottom of the display.

• When calibration of the sensor has been successfully

completed, the instrument will automatically step to the
next sensor to be calibrated and the preceding step will
be repeated.

• When all sensors have been calibrated successfully the

instrument will emit a short beep. The sensor
configuration will be shown on the display and sensor
identifiers will flash to warn of marginal calibration.
Marginal calibration will occur if the sensor full span
value is less than 70% of the applied gas concentration.
A marginal sensor calibration may be an early warning
sign that the sensor will soon need to be replaced.

NOTE: If the sensor full span value is less than 50% of

the applied gas value, the sensor will fail
calibration and the instrument will immediately
return to the real time READING mode. When
failed calibration or low sensitivity is indicated,
verify that the calibration cylinder has not
emptied or that the cylinder expiration date has
not passed.

• After displaying the sensor configuration, the 

instrument will automatically return to the real time
READING mode.

GO  CAL

PRESS

25  LEL

APPLY

Содержание TMX412

Страница 1: ......

Страница 2: ...rinted 1000 5000 Specifications Subject to Change 1001 Oakdale Road Oakdale PA 15071 1500 412 788 4353 Toll Free 1 800 DETECTS FAX 412 788 8353 Multi Gas Monitor TMX412 Instruction Manual G U A R A N...

Страница 3: ...OUR MISSION Design Manufacture Sell Highest quality products for the preservation of life and property Provide Best customer service available...

Страница 4: ...uction Manual it will provide you with years of reliable monitoring I am most concerned that you be pleased with the performance of your TMX412 in the months and years ahead I urge you to call us with...

Страница 5: ...en mixtures only CAUTION High Over Range OR combustible gas readings may indicate an explosive concentration of combustible gas Multi Gas Monitor TMX412 1 WARNINGS AND CAUTIONARY STATEMENTS 3 2 UNPACK...

Страница 6: ...sociation has assessed only the combustible gas portion of this instrument for performance Mine Safety and Health Administration MSHA Workcover Authority NSW Australia Department of Mineral Resource N...

Страница 7: ...ygen Sensor Display Battery Condition Display E Backlight Key Toxic Sensor Port 2 Combustible Gas Sensor Port Combustible Sensor Display Toxic Sensor Display Unmarked keys allow alarm and calibration...

Страница 8: ...splay the type of sensor in the position in which it is installed in the instrument HOLD RELEASE 8 4 INSTRUMENT OPERATION 4 1 CHARGING THE BATTERY CAUTION If the TMX412 is using the nickel cadmium bat...

Страница 9: ...m the combustible gas sensor to prevent damage due to the high level of combustible gas O 0 21 0 OR 10 4 4 3 ZERO This mode allows the user to zero the instrument and calibrate all installed sensors R...

Страница 10: ...nstrument in clean air is preferred provided that there is no trace of toxic or combustible gas that would cause the instrument to respond If the air purity is uncertain use a cylinder of zero grade a...

Страница 11: ...to CAL will scroll across the bottom of the display Press E to continue calibrating the remaining sensors Ignoring this message will result in the instrument returning to the normal operating mode in...

Страница 12: ...ll sensors The display will show the first span value eg 25 LEL along with the prompt PRESS FOR NEXT E TO SELECT Press E to select the value you wish to change The display will flash the current span...

Страница 13: ...HTED AVERAGE The accumulated gas exposure averaged over a predetermined time typically eight hours 7 1 8 STEL SHORT TERM EXPOSURE LIMIT The accumulated gas exposure value averaged for the proceeding f...

Страница 14: ...When changing lithium battery cells replace the cells within the lithium battery pack in a non hazardous location Replace the lithium cells with three Duracel DL123A or three Panasonic CR123A or thre...

Страница 15: ...tely install the sensor in the instrument Never apply pressure to the area inside of the black O Ring seal at the top of the sensor 8 5 2 COMBUSTIBLE SENSOR REMOVAL REPLACEMENT To remove the combustib...

Страница 16: ...24 25...

Страница 17: ...x 25 SST 2 34 1704 2219 Screw FHCR 4 40 x 38 SST 3 35 1704 9736 Conductive Tape RF Screen 1 36 1705 0776 Circuit Board Support 2 37 1704 3308 Battery Spacer 1 38 1704 4066 Data Port Plug 1 39 1704 676...

Страница 18: ...ompact Charger Australian Version 1810 2231 230 VAC Two Unit Compact Charger 1810 1972 230 VAC One Unit Dual Rate Charger 1810 2145 12 VDC One Unit Compact Charger 1810 1915 12 VDC Two Unit Dual Rate...

Страница 19: ...RIGINAL PURCHASE PRICE OF THE NON CONFORMING GOODS IN NO EVENT WILL INDUSTRIAL SCIENTIFIC BE LIABLE FOR ANY OTHER SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES INCLUDING LOSS OF PROFIT OR LOSS OF USE AR...

Отзывы: