background image

17

BATTERY

PRESS

NCD/LI

PRESS

• Press (E) to enter the instantaneous alarms function. The

display will show the first alarm to be set along with the
scrolling prompt PRESS (+) FOR NEXT (E) TO
SELECT. The alarm type will be indicated on the display
as either high (H) or low (L).

• Press (E) to select the desired alarm to be changed. The

display will flash the current alarm value along with the
prompt PRESS (+) OR (-) TO SET (E) TO SELECT.

• Press the (+) and (-) keys to set the desired alarm value

and (E) to enter the value into memory.

• Press MODE to return to the INSTANT function.

• Press (+) to step to the TWA and STEL alarm functions.

The TWA and STEL alarm values are set as 
previously described.

6.3   B

ATTERY

The BATTERY option allows the user to select the type of
battery installed in the unit. The setting configures the
battery indicator to accurately display the discharge rate of
the selected battery type. 

Change the battery setting as follows:

1. From Setup, access the BATTERY option by pressing

the key until “BATTERY” appears on the
display and then press the E key.

2. Press the key to toggle between battery types.

The display will show “NCD/LI” for Ni-Cad or lithium
batteries and “ALKALNE” for alkaline batteries.

3. Press the E key to select battery type shown.

4. Press the Mode key twice to return to Reading Mode for

normal operation of the instrument.

10L  LEL

PRESS

16

LEL

PRESS

25 

LEL

PRESS

• Press (+) to toggle between LEL and CH

4

combustible

sensor span ranges.

• Press MODE to enter the span gas values for all sensors.

The display will show the first span value, eg. 25 LEL,
along with the prompt PRESS (+) FOR NEXT (E) TO
SELECT.

• Press (E) to select the value you wish to change. The

display will flash the current span value and will scroll
the prompt PRESS (+) or (-) TO SET (E) TO ENTER.

• Press the (+) and (-) minus keys to set the desired span

value and (E) to enter the value into memory. Once the
value has been entered into memory, it will become the
standard span gas value used during instrument
calibration.

• Press the MODE switch to return to the SETSPAN

function.

6.2   A

LARMS

The ALARMS function allows you to set the values for
the HI and LOW alarms for each installed sensor. Default
alarm settings for each gas are listed in Section 11. If the
hygiene/datalogging option is installed, you will also be
able to set the STEL and TWA alarm values.

• Press (+) to step from the SETSPAN function to the

ALARMS function.

• Press (E) to enter the ALARMS function. The LATCH

function will lock the alarm indicators on when a
monitored gas concentration reaches the high alarm
setpoint. The alarm will reset after the gas concentration
has fallen below the alarm setpoint and the user has
pressed the (E) key. The display will indicate LATCH
along with the scrolling prompt PRESS (+) FOR NEXT
(E) to SELECT.

• Press (E) to select the LATCH function. The display will

indicate LATCH along with the prompt PRESS (+) FOR
NEXT (E) TO SELECT.

• Press (E) to set the TMX412 high alarm latch as ON or

OFF. Press MODE to return to the LATCH screen.

• Press (+) to step to the INSTANT function and set

instantaneous alarm values. The display will indicate
INSTANT along with the scrolling prompt PRESS (+)
FOR NEXT (E) TO SELECT. 

ALARMS

PRESS

LATCH

LATCH

PRESS

OFF

PRESS

INSTANT

PRESS

Содержание TMX412

Страница 1: ......

Страница 2: ...rinted 1000 5000 Specifications Subject to Change 1001 Oakdale Road Oakdale PA 15071 1500 412 788 4353 Toll Free 1 800 DETECTS FAX 412 788 8353 Multi Gas Monitor TMX412 Instruction Manual G U A R A N...

Страница 3: ...OUR MISSION Design Manufacture Sell Highest quality products for the preservation of life and property Provide Best customer service available...

Страница 4: ...uction Manual it will provide you with years of reliable monitoring I am most concerned that you be pleased with the performance of your TMX412 in the months and years ahead I urge you to call us with...

Страница 5: ...en mixtures only CAUTION High Over Range OR combustible gas readings may indicate an explosive concentration of combustible gas Multi Gas Monitor TMX412 1 WARNINGS AND CAUTIONARY STATEMENTS 3 2 UNPACK...

Страница 6: ...sociation has assessed only the combustible gas portion of this instrument for performance Mine Safety and Health Administration MSHA Workcover Authority NSW Australia Department of Mineral Resource N...

Страница 7: ...ygen Sensor Display Battery Condition Display E Backlight Key Toxic Sensor Port 2 Combustible Gas Sensor Port Combustible Sensor Display Toxic Sensor Display Unmarked keys allow alarm and calibration...

Страница 8: ...splay the type of sensor in the position in which it is installed in the instrument HOLD RELEASE 8 4 INSTRUMENT OPERATION 4 1 CHARGING THE BATTERY CAUTION If the TMX412 is using the nickel cadmium bat...

Страница 9: ...m the combustible gas sensor to prevent damage due to the high level of combustible gas O 0 21 0 OR 10 4 4 3 ZERO This mode allows the user to zero the instrument and calibrate all installed sensors R...

Страница 10: ...nstrument in clean air is preferred provided that there is no trace of toxic or combustible gas that would cause the instrument to respond If the air purity is uncertain use a cylinder of zero grade a...

Страница 11: ...to CAL will scroll across the bottom of the display Press E to continue calibrating the remaining sensors Ignoring this message will result in the instrument returning to the normal operating mode in...

Страница 12: ...ll sensors The display will show the first span value eg 25 LEL along with the prompt PRESS FOR NEXT E TO SELECT Press E to select the value you wish to change The display will flash the current span...

Страница 13: ...HTED AVERAGE The accumulated gas exposure averaged over a predetermined time typically eight hours 7 1 8 STEL SHORT TERM EXPOSURE LIMIT The accumulated gas exposure value averaged for the proceeding f...

Страница 14: ...When changing lithium battery cells replace the cells within the lithium battery pack in a non hazardous location Replace the lithium cells with three Duracel DL123A or three Panasonic CR123A or thre...

Страница 15: ...tely install the sensor in the instrument Never apply pressure to the area inside of the black O Ring seal at the top of the sensor 8 5 2 COMBUSTIBLE SENSOR REMOVAL REPLACEMENT To remove the combustib...

Страница 16: ...24 25...

Страница 17: ...x 25 SST 2 34 1704 2219 Screw FHCR 4 40 x 38 SST 3 35 1704 9736 Conductive Tape RF Screen 1 36 1705 0776 Circuit Board Support 2 37 1704 3308 Battery Spacer 1 38 1704 4066 Data Port Plug 1 39 1704 676...

Страница 18: ...ompact Charger Australian Version 1810 2231 230 VAC Two Unit Compact Charger 1810 1972 230 VAC One Unit Dual Rate Charger 1810 2145 12 VDC One Unit Compact Charger 1810 1915 12 VDC Two Unit Dual Rate...

Страница 19: ...RIGINAL PURCHASE PRICE OF THE NON CONFORMING GOODS IN NO EVENT WILL INDUSTRIAL SCIENTIFIC BE LIABLE FOR ANY OTHER SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES INCLUDING LOSS OF PROFIT OR LOSS OF USE AR...

Отзывы: