background image

13

Industrial Scientific further recommends that a full instru-
ment be performed using a certified concentration(s) of
calibration gas(es) monthly to ensure maximum accuracy.

If an instrument fails to operate properly following any
functional “bump” test, a full instrument calibration
should be performed prior to use.

Calibration will be most accurate when the instrument has
been in a stable temperature environment for at least one
hour prior to calibrating.

NOTE: Instrument zero and oxygen span calibration

should be performed in clean air containing
20.95% (21%) oxygen. If you are measuring a
known combustible gas, use a known %LEL
concentration of that gas for span calibration. For
general combustible gas measurement, Industrial
Scientific Corporation recommends calibrating to
pentane in the 15-50% LEL range. 
The measured LEL concentration of gases other
than the calibration gas may not correspond on a
one-to-one basis with the monitor reading.
Calibration gas flow rate should be between 0.5
and 1.0 liters per minute. Apply CL

2

, at a 1.0 liter

per minute flow rate. Always use teflon or teflon
lined tubing when calibrating Cl

2

or NO

2

sensors.

The TMX412 utilizes a one-button calibration system.
When using multi-component gas cylinders, a full
instrument span calibration can be performed in a 
single step. Multi-component calibration gas cylinders are
available for combustible gas/carbon monoxide/hydrogen
sulfide configurations.

To calibrate the TMX412:
• From the normal READING mode, press the MODE

switch to access the ZERO operating mode.

• Press the (E) key to start the instrument zeroing process.

The instrument display will indicate ZEROING.

• When zeroing is complete the instrument display will

indicate O2 CAL and will show the current full span
value of the oxygen sensor. The oxygen sensor will be
spanned to 21.0 in approximately 30 seconds. Oxygen
calibration may be aborted at this time by pressing the
MODE switch.

NOTE: Zeroing the instrument in clean air is preferred,

provided that there is no trace of toxic or
combustible gas that would cause the instrument
to respond. If the air purity is uncertain, use a
cylinder of zero grade air to zero the instrument
and span the oxygen sensor.

ZERO

PRESS

ZEROING

21.0

02  CAL

12

BATTERY

FAIL

To clear the combustible gas over-range alarm:

• Exit the hazardous area immediately.
• Press the enter (E) key in clean air.

NOTE: When the enter (E) key is pressed, the combustible

gas level must be less than 100% of LEL or 5%
CH

4

to clear the combustible gas over-range

condition.

4.5.5

LOW BATTERY WARNING

When approximately 1 hour of instrument operation
energy remains in the battery, the instrument emits a short
beep every 15 seconds. In addition, the battery indicator
will be replaced by a flashing “B” on the display.

NOTE: The length of warning time will increase when

there is no combustible gas sensor installed in the
instrument.

4.5.6

BATTERY FAILURE

When the battery has insufficient charge to operate the
instrument, “BATTERY FAIL” is displayed. The
instrument stops monitoring and emits a short beep once
every second.

When “BATTERY FAIL” appears, turn off the instrument
immediately. Recharge or replace the battery.

4.5.7

FAULT INDICATION

The TMX412 will emit a short beep approximately once a
second if a newly installed sensor does not agree with the
valid sensor type for that position when the instrument is
turned on. The corresponding display position will be
blank. This fault indication occurs when a toxic or
combustible sensor becomes disconnected or a
combustible fault is detected during normal operation.
Sensor types become valid only after a successful
calibration has been completed.

5.

C

ALIBRATING THE

TMX412

The TMX412 is a potential life saving device.
Recognizing this fact, Industrial Scientific Corporation
recommends that a functional (“bump”) test be performed
on every instrument prior to each days use. A functional
test is defined as a brief exposure of the monitor to a
known concentration of gas(es) for the purpose of
verifying sensor and alarm operation and is not intended to
be a measure of the accuracy of the instrument.

Содержание TMX412

Страница 1: ......

Страница 2: ...rinted 1000 5000 Specifications Subject to Change 1001 Oakdale Road Oakdale PA 15071 1500 412 788 4353 Toll Free 1 800 DETECTS FAX 412 788 8353 Multi Gas Monitor TMX412 Instruction Manual G U A R A N...

Страница 3: ...OUR MISSION Design Manufacture Sell Highest quality products for the preservation of life and property Provide Best customer service available...

Страница 4: ...uction Manual it will provide you with years of reliable monitoring I am most concerned that you be pleased with the performance of your TMX412 in the months and years ahead I urge you to call us with...

Страница 5: ...en mixtures only CAUTION High Over Range OR combustible gas readings may indicate an explosive concentration of combustible gas Multi Gas Monitor TMX412 1 WARNINGS AND CAUTIONARY STATEMENTS 3 2 UNPACK...

Страница 6: ...sociation has assessed only the combustible gas portion of this instrument for performance Mine Safety and Health Administration MSHA Workcover Authority NSW Australia Department of Mineral Resource N...

Страница 7: ...ygen Sensor Display Battery Condition Display E Backlight Key Toxic Sensor Port 2 Combustible Gas Sensor Port Combustible Sensor Display Toxic Sensor Display Unmarked keys allow alarm and calibration...

Страница 8: ...splay the type of sensor in the position in which it is installed in the instrument HOLD RELEASE 8 4 INSTRUMENT OPERATION 4 1 CHARGING THE BATTERY CAUTION If the TMX412 is using the nickel cadmium bat...

Страница 9: ...m the combustible gas sensor to prevent damage due to the high level of combustible gas O 0 21 0 OR 10 4 4 3 ZERO This mode allows the user to zero the instrument and calibrate all installed sensors R...

Страница 10: ...nstrument in clean air is preferred provided that there is no trace of toxic or combustible gas that would cause the instrument to respond If the air purity is uncertain use a cylinder of zero grade a...

Страница 11: ...to CAL will scroll across the bottom of the display Press E to continue calibrating the remaining sensors Ignoring this message will result in the instrument returning to the normal operating mode in...

Страница 12: ...ll sensors The display will show the first span value eg 25 LEL along with the prompt PRESS FOR NEXT E TO SELECT Press E to select the value you wish to change The display will flash the current span...

Страница 13: ...HTED AVERAGE The accumulated gas exposure averaged over a predetermined time typically eight hours 7 1 8 STEL SHORT TERM EXPOSURE LIMIT The accumulated gas exposure value averaged for the proceeding f...

Страница 14: ...When changing lithium battery cells replace the cells within the lithium battery pack in a non hazardous location Replace the lithium cells with three Duracel DL123A or three Panasonic CR123A or thre...

Страница 15: ...tely install the sensor in the instrument Never apply pressure to the area inside of the black O Ring seal at the top of the sensor 8 5 2 COMBUSTIBLE SENSOR REMOVAL REPLACEMENT To remove the combustib...

Страница 16: ...24 25...

Страница 17: ...x 25 SST 2 34 1704 2219 Screw FHCR 4 40 x 38 SST 3 35 1704 9736 Conductive Tape RF Screen 1 36 1705 0776 Circuit Board Support 2 37 1704 3308 Battery Spacer 1 38 1704 4066 Data Port Plug 1 39 1704 676...

Страница 18: ...ompact Charger Australian Version 1810 2231 230 VAC Two Unit Compact Charger 1810 1972 230 VAC One Unit Dual Rate Charger 1810 2145 12 VDC One Unit Compact Charger 1810 1915 12 VDC Two Unit Dual Rate...

Страница 19: ...RIGINAL PURCHASE PRICE OF THE NON CONFORMING GOODS IN NO EVENT WILL INDUSTRIAL SCIENTIFIC BE LIABLE FOR ANY OTHER SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES INCLUDING LOSS OF PROFIT OR LOSS OF USE AR...

Отзывы: