11
Operation-mode bump test
This screen features the status, lock, configuration, and bump test icons,
with the selected setting value displayed in the main area. The technician
can enable or disable this operation mode feature.
When enabled, the instrument operator can bump test the unit from
operation mode.
Values:
0 = disabled ,1 = enabled
Bump test percentage
This screen features the status, configuration, and bump test icons, with
the current setting value displayed in the main area. The technician can set
the percentage of calibration gas to which the unit will respond.
Value range: 50% to 95%
Value increment: 1%
Bump test response-time
This screen features the status, clock, configuration, and bump test icons,
with the current setting in seconds displayed in the main area. A sensor
passes a bump test when it senses the specified percentage of calibration
gas within the specified response time setting.
Value range: 30 to 120 seconds
Value increment: 5 second
Alarm latch
This screen features the status, alarm, lock, and configuration icons, with
the current setting displayed in the main area. The technician can enable
or disable this operation-mode feature.
When disabled, a unit in alarm will turn off its alarm when the gas reading
is no longer at the alarm-producing concentration.
When enabled, a unit in alarm will remain in alarm until it is manually reset.
The instrument operator can reset a latched alarm from gas normal reading
screen by click button.
Values:
0 = disabled ,1 = enabled