12
Vibration alarm
This screen features the status, alarm, “VIb”, and configuration icons, with
the selected setting value displayed in the main area. When enabled, the
vibrating alarm will be activated when the unit is in alarm.
Values: 0 = disabled ,1 = enabled
Operation-mode zero
This screen features the status, lock, zero, and configuration icons. The
technician can enable or disable this operation-mode option.
When enabled, the instrument operator can zero the unit from operation
mode.
Values: 0 = disabled ,1 = enabled
Operation-mode calibration
This screen features the status, lock, configuration, and calibration icons.
The technician can enable or disable this operation-mode option.
When enabled, the instrument operator can calibrate the unit from
operation mode.
Values:0 = disabled ,1 = enabled
Security code
This screen features the status, lock, and configuration icons, with the
current security code displayed in the main area. The security code
controls access to a unit's configuration mode and the ability to power off a
unit that is configured for always-on operation.
If the security code is set at 000, entry to configuration mode is not
security-code protected, and an always-on unit can be powered off without
a security code. Any other value will enable the security code.
Value range: 000 to 999
Value increment: 1
Always-on
This screen features the status, lock, configuration, and the battery icon.
The technician can enable or disable this feature.
When enabled, the entry of the unit's security code (if the security code is
not 000) will be required to complete the shutdown process.
Values:
0 = disabled ,1 = enabled