TROLL Link 100 Setup
This section describes configuring a TROLL Link 100 System. For TROLL Link 101 or
201 Systems, skip to the section
TROLL Link 101 and 201 Setup
.
You MUST configure and test your TROLL Link Telemetry System
in your home or office prior to installing it in the field.
Changing Modbus Communication Settings in Win-Situ 5
Complete the following steps with the instrument connected to Win-Situ 5 Software.
Change Modbus communication settings for each instrument.
1. Click the
Device Setup
tab.
2. Click the
Modbus Setup...
button.
3. Click the
Baud
drop-down menu and select the correct baud rate for your cable
length based on the on-screen instructions. The default value is 19200.
4. Click the
Parity Bits
drop-down menu and select
None
.
5. Click the
End of Session Timeout (ms)
box and enter
60000=60secs
.
6. Click the
Mode
drop-down menu and select
Modbus-ASCII
.