Option
Description
Send current readings using
SMS Messages
Sends an SMS (text) message that has been
configured in the
Messages Tab
. You must select
the type of TROLL connected to the TROLL Link
with the radio buttons below.
Wait for a phone call from
Win-Situ 5
Connect to the TROLL Link Telemetry System
using Modem Communications (dial-up).
Wait for an Internet (TCP/IP)
connection from Win-Situ
Connect to the TROLL Link Telemetry System
using IP (internet) Communications.
* If a TROLL 9500 is connected, you must select
Send current readings using SMS
Messages
. You will not be able to download data from a TROLL 9500 in dial-up or TCP/IP
modes.
If you select
Wait for an Internet (TCP/IP) connection from Win-Situ
, you must
determine the TROLL Link Modem IP and Port number. These are displayed on the
Summary Tab as
Modem IP
and
Port
. Write these numbers in the space provided
below.
Modem IP: _________._________._________._________
Port: _________ (3001 is default)
3. Select what type of instrument(s) are connected to the TROLL Link. Choose the
communication protocol. The protocol for Level TROLL and Aqua TROLL
Instruments is ASCII.
4. Leave the Compression Modbus Messages box unchecked.
5. Select an Auto Reset Modem Interval. The reset interval determines how often the
modem power supply cycles (turns off and back on). Four hours is typical.
6. Enter a 10-character access password if desired. Write the password in the space
provided below. Click the
Set Pswd
button.
TROLL Link Access Password: ____________________