ISOCON-6 USERS GUIDE
8. SPECIFICATIONS ( @ 25°C):
Operating Temperature: 0 to 55 °C
Operating Altitude: Sea Level to 2000m
Humidity: 0-90% RH
Power Requirements:
DC Supply: 12 to 36Vdc
AC Supply: 12 to 32Vac
Current Consumption: 55mA @ 24Vdc (20mA in & out)
85mA @ 24Vdc (maximum load, tx supply)
200mA @ 12Vdc (maximum load, tx supply)
260mA for 50ms on 24Vdc power up
Transmitter Power Supply: 22V to 29V @ up to 24mA
Dependant on supply voltage and load
Calibration accuracy: ±0.05% full scale
Linearity: ±0.05% full scale
Temperature Stability: 50ppm / °C
Input Impedance:
Current Input: 15 ohms
Voltage Input: 1 Mohm
Millivolt Input: Greater Than 10 Mohm
Thermocouple Burn Out Current: 500nA Nominal
Cold junction compensation accuracy: ±0.5°C over operating range
Maximum Voltage Output: 11.5 V into a minimum of 7Kohm
Maximum Current Output: 23.0 mA into a maximum of 1Kohm
Time Response: (90% of step change): 50ms ± 10ms
Unit has full 3 port Isolation to 1kV between Power Supply, Input and Output.
The unit can also withstand transients of 2.5kV for 50 μsecs.
Dimensions: 114.5 mm x 99mm x 12.5mm (H x D x W)
Mounting: DIN Rail TS35
Connections: Screw Clamp with pressure plate
Conductor Size: 0.5 to 4.0 mm
Insulation Stripping: 12 mm
Maximum Terminal Torque: 0.4 Nm
Weight Approx. 95g
EMC Emissions:
EN50081-1:1992
EN50081-2:1993
EMC Immunity: EN50082-1:1997
LVD Standards: EN61010-1:1993
Installation Category (IEC 664): II
Pollution Degree (EN61010-1:1993): 2
Equipment Class (IEC 536): II
ISOCON-6
24V AC or DC POWERED ISOLATING SIGNAL CONVERTER