3.7. Exchange of the Probe Head of the VARIO
Xtrem/LD
At the SONO-VARIO
Xtrem
and VARIO
LD
not only the ceramic plate but the whole metal/ceramic wear-
resistant probe head can be exchanged.
This is how easy it is to exchange the wear-resistant probe head:
- Loosen the 4 fastening screws (look for the right sequence of washer and gasket rings).
- Lift off the probe head carefully so that the robust spring contacts in the interior are canted as
little as possible.
- Clean the surface inside the probe body for the O-Ring.
- Place the probe head so that the two spring contacts are inserted in the contact bushings.
- Screw the 4 fastening screws back on.
It is to consider, that the 4 screws “find” the 4 holes in the
green epoxy plate inside the probe.