35/44
3.3. Mounting
in curved Surfaces
In order to prevent the probe head from projecting and interfering with wipers, the centre of the
probe head should be flush with the radius of the wall. The ceramic must be laterally aligned to
the rotational axis as the gap towards the wipers is smallest in this position.
In order to ensure that the probe head is completely covered with material, the probe is best
positioned near the base. In order to prevent water from accumulating above the sensor head,
which could falsify the measurement, an installation angle of approximately 30° above the base
centre is recommended.
3.4. Funnel shape for higher material depth
If a too quick material flow leads to a too low material height above the probes surface, than a guided
funnel shape plate arrangement under a vibration chute could solve the problem. The plates are
ideally coated with PTFE for a better adherence behaviour.