![IMKO SONO-VARIO LD Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/imko/sono-vario-ld/sono-vario-ld_user-manual_2047643023.webp)
23/44
1.8. Connection of the RS485 to the SM-USB Module from IMKO
The SM-USB provides the ability to connect a SONO probe either to the standard RS485 interface or
optionally to the IMP-Bus from IMKO, which enables the download of a new firmware to the SONO
probe. Both connector ports are shown in the drawing below.
The SM-USB is signalling the status of power supply and the transmission signals with 4 LED´s. When
using a dual-USB connector on the PC, it is possible to use the power supply for the SONO probe
directly from the USB port of the PC without the use of the external AC adapter.
How to start with the USB-Module SM-USB from IMKO
Install USB-Driver from USB-Stick.
Connect the SM-USB to the USB-Port of the PC and the installation will be accomplished
automatically.
Install Software
SONOConfig-SetUp.msi
from USB-Stick.
Connection of the SONO probe to the EX9531 via RS485A, RS485B and 0V.
Check the setting of the COM-Ports in the Device-Manager und setup the specific COM-Port with
the Baudrate of 9600 Baud in SONO-CONFIG with the button "
Bus
" and "
Configuration
" (COM1-
COM15 is possible).
Start
“
Scan probes
” in SONOConfig.
The SONO probe logs
in the window „
Probe List
“ after max. 30 seconds with its serial number.