3.1. Further Assembly Instructions
With the universal holder, deliverable by Imko, the SONO-VARIO can be placed above a
conveyor belt. Particularly with inhomogeneous or loose sand, the material can be constantly
condensed, which leads to higher measurement results.
The installation locations may not be situated beneath the inlets for additives.
In case of an uneven base, the probe must be installed at the highest point of the base. No
water may accumulate at the probe head as this could falsify the measurement.
Areas with strong turbulences are not ideal for the installation. There should be a
continuous material flow above the probe head.
The stirring movement of blades should be conducted without gap above the probe head.
The probe should not be installed in the direct vicinity of electrical disturbing sources such
as motors.
In case of curved installation surfaces in containers, the centre of the probe head should be
flush with the radius of the container wall without disturbing the radial material flow in the
container. The probe may not project and come in contact with blades or wipers.
Attention! Risk of Breakage!
The probe head is made of special steel and a very wear-resistant ceramic in order to warrant for
a long life-span of the probe. In spite of the robust and wear-resistant construction, the ceramic
plate may not be exposed to any blows as ceramic is prone to breakage.
In case of welding work at the plant, all probes must be completely electrically
disconnected.
Any damage caused by faulty installation is not covered by the warranty!