After the spectrometer is calibrated and has been reinstalled to your system, perform a spectral 
calibration (iQ-LED calibration) and check if this intensity value is still correct. If it is not correct, 
you have to perform a lux calibration.  

4.3.3  iQ-LED calibration 

The individual LED lights of the iQ-LED inside the CAL1 depends on many different types and 
wavelengths. Some LEDs will change their intensity level and peak wavelength slightly in the 
first 500-600 working hours because of a burn-in effect.  

The LEDs will also degrade in intensity during their lifetime. To make sure that all 
measurements including the auto-generated illuminants and the standard illuminants, are 
correct, you have to perform a spectral calibration regularly.  

You must also consider the degradation of the LED when saving self-defined presets. If you 
save a preset with LED channels that uses its maximum intensity, the possibility exists that this 
intensity cannot be reached after the burn-in time or the long-time degradation of the LED. In 
this case, you will get a warning message from the iQ-LED control software. 

During the first 500-600 working hours, we recommend performing a spectral calibration every 
50 operating hours. 

After the first 500-600 operating hours, a calibration of every 150 working hours suffices. 

Other factors that indicate the need for a spectral calibration: unsatisfactory illuminant 
generation, the aberration of the intensity values, or a spectral curve that does not fit with the 
predefined standard illuminants of the corresponding preset.  

  the spectrometer works correctly 

  the spectrometer settings are correct 

  all LED channels work correctly 

  the dark measurement is correct 

  your measurement environment is correct 

  your ambient temperature is correct 

 

How to perform the spectral calibration is described in the iQ-LED control software manual. 

4.4  Low intensity use 

When using your system with very low intensity, the spectral measurement values will start to 
fluctuate. The lower the intensity, the higher the fluctuation. The generated light is still stable up 
to a certain point. The fluctuation of the values is caused by the noise of the spectral 
measurement of the internal spectrometer. The light intensity will continue to get lower when the 
influence of the noise continues to get higher. When using standard illuminants with an intensity 
lower than 25 lx, it will no longer be possible to get a correct value.   

 

 

 

Содержание CAL1

Страница 1: ...Image Engineering GmbH Co KG Im Gleisdreieck 5 50169 Kerpen Horrem Germany T 49 2273 99991 0 F 49 2273 99991 10 www image engineering com CAL1 User Manual 18 December 2018 ...

Страница 2: ...TRUCTIONS HARDWARE 5 3 1 Overview display and ports 5 3 2 Connecting the hardware 6 3 3 Camera positioning 7 4 OPERATING INSTRUCTIONS SOFTWARE 7 4 1 Requirements 7 4 2 Software installation 7 4 3 Starting the system 7 4 3 1 Spectrometer settings 7 4 3 2 Spectrometer calibration 7 4 3 3 iQ LED calibration 8 4 4 Low intensity use 8 5 ADDITIONAL INFORMATION 9 5 1 Maintenance 9 5 2 Care instructions 9...

Страница 3: ...ronment without any interfering light The optimal ambient temperature range is 22 to 26 degrees Celsius The maximum ambient temperature range is 18 to 28 degree Celsius The optimal system temperature range displayed in the software user interface is between 35 and 50 degrees Celsius The system has internal temperature management if there is any error regarding the internal temperature you will get...

Страница 4: ... team or when connected to the power supply 2 GETTING STARTED 2 1 Scope of delivery integrating sphere spectrometer built in power cord USB cable control software calibration protocol Optional equipment iQ Align for CAL1 for a quick and easy camera alignment EX2 spectrometer for external measurements iQ Trigger The iQ Trigger is a mechanical finger that can press the release button within 25 ms Wh...

Страница 5: ... storage of the illuminants with the play and stop button you can start and stop a saved light sequence with different illuminants it is possible to save one sequence on the device with the power button you can turn on and off the light there are three pre stored illuminants on your device the intensity of each illuminant is shown in the acceptance protocol of your device o 1 illuminant A default ...

Страница 6: ...d to synchronize your test setup For example with an iQ Trigger See 2 1 optional equipment 3 2 Connecting the hardware 1 Connect the power cord to the power supply on the back of CAL1 2 Connect the USB cable to the CAL1 and your PC 3 Turn on the CAL1 the power switch is located beside the power supply 4 Connect the spectrometer to the PC using the USB cable The system will install the spectrometer...

Страница 7: ...sktop Follow the iQ LED software manual to control the CAL1 NOTE The iQ LED devices can only operate with high precision when setup and calibration are performed correctly Consult the iQ LED software manual for a comprehensive description and read it carefully 4 3 1 Spectrometer settings The iQ LED software see iQ LED software manual automatically generates the best spectrometer settings for you l...

Страница 8: ... LED control software During the first 500 600 working hours we recommend performing a spectral calibration every 50 operating hours After the first 500 600 operating hours a calibration of every 150 working hours suffices Other factors that indicate the need for a spectral calibration unsatisfactory illuminant generation the aberration of the intensity values or a spectral curve that does not fit...

Страница 9: ...d to bridge the time gap while your spectrometer is recalibrated Please contact us for conditions and procedure 5 2 Care instructions Do not touch scratch or pollute the diffusor If there is any dust on the diffusor clean it with an air blower Do not remove the fiber from the spectrometer Otherwise the calibration is invalid and the spectrometer has to be recalibrated Only store and transport the ...

Страница 10: ...ical data sheet Image Engineering CAL1 Seite 10 von 10 6 TECHNICAL DATA SHEET See annex for the technical data sheet It can also be downloaded from the website of Image Engineering www image engineering com ...

Отзывы: