5
TORQUE MEASUREMENT
Program
1
Modes
With the power OFF, press and hold
the
M
button and then press and
release (do not hold) the
P
button.
When
1
appears on the left side of
the screen, press
–
or
+/C
to cycle:
P
Peak mode
captures peak torque
A
Peak mode w/auto zero reset
captures peak torque and resets to
zero at programmed interval
t
Real time mode
displays transient
torque values and peak is not held.
Press and release the
M
button to
advance to the next setting.
Program
2
Auto zero reset duration.
If peak mode w/auto zero reset (
1 A
)
has not been previously selected,
press and release the
M
button to
skip this step.
If peak mode w/auto zero reset has
been selected, press
–
or
+/C
to
change the duration a peak value
remains on screen before it resets to
zero 1~10 seconds, (factory default is
2 seconds). Press and release the
M
button to advance to the next setting.
Program
3
Audible beep
Confirms ready mode and selections,
alerts to torque presets and limits.
Press
–
or
+/C
to turn ON or OFF.
Press and release the
M
button to
advance to the next setting.
1
CYCLE WITH – OR +/C KEYS
INITIAL PROGRAMMING
7
p A
t
2
CYCLE WITH – OR +/C KEYS
3
1 2... 10
o
n
off
CYCLE WITH – OR +/C KEYS