background image

TORQUE MEASUREMENT

Complete color 

catalog available.

Call 800-373-9989

or E-mail us.

Specifications subject to change without notice.

Imada, Inc.

®

3100 Dundee Rd., Suite 707, Northbrook, IL 60062 USA
Telephone: (847) 562-0834     Fax: (847) 562-0839
www.imada.com   E-mail: [email protected]
ISO 9001 & ISO/IEC 17025 Accredited

LIMITED WARRANTY

TWO-YEAR LIMITED WARRANTY AND LIMITATION OF LIABILITY:  

Imada, Inc., for itself and

its related companies (hereinafter described collectively as “Imada”) warrants that all Imada prod-
ucts sold, except for those specific products identified below, will be free of defects in materials and
workmanship for a period of two (2) years from date of delivery to end-user.  The foregoing two (2)
year warranty shall not be extended or changed by Imada furnishing any replacements, additions,
attachments, accessories or repairs to the products subsequent to the date of delivery or accept-
ance. The foregoing warranty is the sole and exclusive warranty of Imada regarding the products.

ONE-YEAR LIMITED WARRANTY FOR SPECIFIC PRODUCTS: 

Notwithstanding the foregoing,

for Cedar Torque Testers and KTC Torque Wrenches, the limited-warranty period is one (1) year
from date of delivery to end-user.

PRODUCTS EXPRESSLY NOT COVERED BY WARRANTY:

Batteries, cables, attachments, adapters,

and fixtures regardless of the manufacturer of the product, that are sold or provided for by Imada
are not covered by the foregoing warranties. Warranties of the respective manufacturers for such
products shall apply.

DISCLAIMER OF OTHER WARRANTIES: 

Other than the foregoing warranties, there are no

express or implied warranties or any affirmations of fact or promises by Imada with respect to the
products.  Imada disclaims any warranties, express, implied or statutory, not specifically set forth
above. Without limiting the generality of the foregoing, Imada expressly disclaims any implied war-
ranties of merchantability, fitness for any particular purpose, infringement or any representations of
fact or quality not expressly set forth herein.

LIMITATION OF LIABILITY AND REMEDIES:  

Imada's sole responsibility and liability incurred

as a result of the sale and/or use of the products, and the purchaser’s exclusive remedy against
Imada under any warranty shall be limited to the repair or replacement, at Imada's option, of product
components not conforming to the warranty.  The total liability of Imada shall in no event exceed
the amount actually paid to Imada by purchaser with respect to the products.  This limitation of
remedy is intended by the parties to survive even if the remedy is claimed to have failed of its
essential purpose.  Purchaser’s full and complete performance of all obligations of purchaser recited
in this agreement is a condition precedent to Imada's warranty obligations and liabilities herein.

PURCHASER'S DAMAGES AND LIMITATIONS:

In no event shall Imada be liable to purchaser, its

assigns or agents, for economic loss, incidental or consequential damages, in contract or in tort,
including but not limited to, any damages for lost profits, down-time, lost production, failure to meet
purchaser’s sales contracts, or defects in purchaser’s materials or workmanship arising directly or
indirectly from the use of the products.  Imada’s limited warranty is terminated in the event of prod-
uct overload, shock load, misuse, negligence, accident, or if repairs are attempted by the customer or
third-parties without notice to and authorization by Imada.

INSTRUCTIONS FOR OBTAINING WARRANTY REPAIR OR REPLACEMENT: 

Please contact

our customer service department for a return authorization number.  After you have obtained the
authorization number, please return the subject product to Imada with all freight, duty and broker-
age costs prepaid.  You will be notified by Imada regarding the repair or replacement of the prod-
uct.  In some instances, Imada may require proof of purchase.

Содержание GLK060E

Страница 1: ...Models GLK060E GLK250E GLK500E I N S T R U C T I O N M A N U A L 04 18 Compact Ergonomic Digital Torque Screwdrivers...

Страница 2: ...Duration 5 3 Audible Beep 5 4 Counter 6 5 Auto Shutoff Duration 6 6 Change Units 6 GENERAL OPERATION 1 Peak Mode 7 2 Peak Mode with Auto Reset 7 3 RealTime Mode 8 4 Setting Go No Go Presets 8 9 5 Go N...

Страница 3: ...not disassemble or modify the unit 3 INTRODUCTION GLK Digital Torque Screwdrivers combine ergonomic grips digital accuracy and five Go No Go torque presets that alert users with an intermittent signal...

Страница 4: ...ton stores torque values and advances program functions button decreases values C button increases values resets the screen to zero and clears errors LED Indicator signals power on values stored and a...

Страница 5: ...zero reset duration If peak mode w auto zero reset 1 A has not been previously selected press and release the M button to skip this step If peak mode w auto zero reset has been selected press or C to...

Страница 6: ...uto shutoff duration Press or C to cycle 1 10 min utes Factory default is 2 minutes Press and release the M button to advance to the next setting Program 6 Change default unit Press or C to cycle lbf...

Страница 7: ...ton 7 Press and hold the P button for more than two seconds to turn OFF the power 8 2 Peak Mode with Auto Reset Set to peak w auto zero 1 A see page 5 1 Press and release do not hold P to turn on the...

Страница 8: ...GO NO GO Presets Digital torque screwdrivers save five Go No Go presets The presets alert the user with intermittent alerts at 90 of the preset torque value and continuous alerts at 100 and over Pres...

Страница 9: ...other Go No Go presets press and release the M button to advance to presets two through five Check the preset indicator on the left side of the display and fol low steps 3 and 4 6 Press P and hold for...

Страница 10: ...button to select a preset Apply torque gradually and indicators appear as follows INTERMITTENT FLASH AND BEEP AT 90 OF PRESET CONTINUOUS LIGHT AND BEEP AT 100 OF PRESET 5 GO NO GO Preset Operation 9 0...

Страница 11: ...d fas teners between low and high limits in the counterclockwise direction for left hand threads OFF No counter general operation After selecting 4r 0n or 4L 0n press and hold the P button for more th...

Страница 12: ...appears on the left side of the dis play Press the or C buttons to change the high limit M blinks Press and hold the M button until the LED flashes the alert beeps and the display returns to zero Pre...

Страница 13: ...ve the high limit are indi cated by a fast intermittent signal with LED and audible alerts and are not counted Press C to clear the value press again to clear the count To Return to General Operation...

Страница 14: ...gain E2 Turn the power OFF and ON again E3 Requires repair E4 Requires repair E5 Requires repair E6 Turn the power OFF and ON again E7 Turn the power OFF and ON again ERROR CODES When applied torque e...

Страница 15: ...ity Auto Shut OFF Automatic shut off after selectable interval of non use 1 10 min Operating Temp 5 to 40 C Storage Temp 20 to 60 C Battery Lithium coin battery Panasonic CR2354 only Battery Life Appr...

Страница 16: ...expressly disclaims any implied war ranties of merchantability fitness for any particular purpose infringement or any representations of fact or quality not expressly set forth herein LIMITATION OF LI...