M A I N T E N A N C E
Calibration Overview
7/01
MPS-8033
77
C H A P T E R
5
!
Local Operation Optical Power Calibration
The following procedure is for manual calibration of the MPS−8033 optical output power. Remote
calibration is discussed in Remote Operation Optical Power Calibration on page 78. The instrument
utilizes a 3−point calibration to characterize the power response of the source’s laser diode.
Calibrating from the Front Panel
1
Connect the 8033 to an appropriate AC power source. Turn the power on and allow the
8033 to warm up for at least one hour. Enable the laser output by pressing the
(OUTPUT)
ON
switch.
2
Connect the Optical Power Meter to an appropriate AC Power Source. Turn on the power
and configure it for the following:
3
Comply with the warm−up considerations for the power meter. "Zero" the power meter per
the Power Meter’s instructions.
4
Connect the calibration fiber optic patch cord to the MPS−8033.
5
Connect the other end of the patch cord to the Optical Power Meter. Make sure the
connections are clean and tight.
6
Start the calibration by simultaneously pressing
LOCAL/ADR
and
(MODE) CW
. The
display clears for two seconds before the output is turned on and the power is set to the
first calibration value.
7
Using the ADJUST KNOB, change the power until the 8033 display exactly matches the
power displayed on the power meter.
8
Press
SET REFERENCE
to store the first calibration point into memory. The 8033 then
switches to the next calibration value.
9
Use the ADJUST KNOB, change the power until the 8033 display exactly matches the
power displayed on the power meter.
10
Press
SET REFERENCE
to store the second calibration point into memory. The 8033
switches to the final calibration value.
11
Using the ADJUST KNOB, change the power until the 8033 display exactly matches the
power displayed on the power meter.
12
Press
SET REFERENCE
to store the third calibration point into memory. The 8033
internally computes the new calibration factors and returns the device to the same state
as when the calibration was initiated.
The calibration is complete. The new calibration factors are stored in non−volatile RAM for
safe keeping when the power to the device is turned off.
MODE:
Watts (linear)
RANGE:
Auto or mW
FILTER:
Slow
WAVELENGTH:
Equal to the wavelength of 8033 source
(i.e. 1310 nm).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание MPS-8033 Series
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 53: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 103: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 111: ...I N D EX iv MPS 8033 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...