RETCV_0107
45
The IKAMAG RET control-visc /C is maintenace-free. It is sub-
ject only to the natural wear and tear of components and their
statistical failure rate.
When ordering spare parts, please give the manufacturing
number shown on the type plate, the machine type and
the name of the spare part.
Please send in equipment for repair only after it has been cle-
aned and is free from any materials which may constitute a
health hazard. Use only cleansing agents which have been
approved by IKA to clean IKA devices. To remove use:
Dyes
isopropyl alcohol
Construction materials water containing tenside / isopropyl alcohol
Cosmetics
water containing tenside / isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
For materials which are not listed, please request information
from IKA. Wear the proper protective gloves during cleaning of
the devices.
Electrical devices may not be placed in the cleansing agent for
the purpose of cleaning.
Before using another than the recommended method for clea-
ning or decontamination, the user must ascertain with the
manufacturer that this method does not destroy the instru-
ment.
Send in the case of service the equipment back in the origina
packaging zurück. Storage packing is not sufficient for the back
dispatch. Use additionally a suitable transpertation packing.
16.1
RET control-visc C Technical Enamel
The technical enamel makes the heating plate easier to care
fort and more resistant to acids and bases. Because of it,
however, theheating plate is also more susceptible to extreme
fluctuations in temperature and the force of impact. This can
result in cracks forming or the coating flaking off.
Make certain that the bottom of the placing vessel is even,
clean and dry. The bottom of the placing vessel must not have
any sharp grooves, sides or edges. Remove residues of bases
and immedialety. We recommend most strongly that you clean
the heating plate regularly.
16.1.1
Protective conductor RET control-visc C
On the lower surface of the heating plate is a metallically bright
place. It can be used as a test point for the protective conduc-
tor.
Construction in accordance with the foolwing safety
standards
EN 61010-1
UL 3101-1
CAN/CSA C22.2 (1010-1)
EN 61010-2-10
Construction in accordance with the foolwing EMC stan-
dards
EN 61326-1
Associated EU guidlines
EMC-guidlines:
89/336/EWG
Machine guidelines:
73/023/EWG
F-F
Z 1:2
F
F
Prüfstelle für Schutzleiterprüfung der Heizplatte
bei RET cv C mit isoliertem Prüfstift Ømax. 5mm,
ca. 70mm lang, von unten durch Bohrung Ø 7mm
Maintenance and cleaning
16
Associated standards
and regulations
17
Inspection station for protectivegrounding examination
by RET cv C with isolated test pin Ømax. 5mm,
ca. 70mm long, from down duch drilling Ø 7mm
Содержание WERKE IKAMAG RET control-visc
Страница 105: ...104 RETCV_0107 ...
Страница 106: ...105 RETCV_0107 ...