17
032008
Commissioning
Stirring
Heating
Connecting
external
thermometers
(direct temper-
ature control in
the media)
REO basic
RET basic
safety control
Move device switch (A) to the OFF position
Insert the mains power cable into the power socket (K)
Move device switch (A) to the ON position
Operating mode B is preset
Operating mode A is preset (see “Operating modes”)
Operation
Adjust the motor speed using the rotary knob (D)
The value selected will be shown on the display (F)
Set the start point for the agitation function by pressing the rotary knob (D)
The displayed value will flash until the desired motor speed is reached
i
Ensure that the protective film is removed from the base plate before use!
Move device switch (A) to the OFF position
Detach contact plug (L)
Attach a DIN 12878 (Class 2) compliant safety contact thermometer or a
PT 1000 temperature sensor to the socket (L)
Move device switch (A) to the ON position
Temperature sensor PT 1000 The actual temperature for the temperature sensor shown on
display (E) will correspond to the temperature of the media.
The LED "external temperature sensor" (H) will be lit
Contact thermometer e.g. ETS-D5 Follow the operating instructions for the contact thermometer
The LED "decimal point for the temperature display" (J) will be lit
When a contact thermometer is connected, the display (E)
will only show the set-point temperature that has been set
i
i
Set the safe temperature limit (see "Setting the safe temperature limit for RET")
Adjust the set-point temperature using the rotary knob (C)
The value selected will be shown on the display (E)
Set the temperature control mode (see “Setting the temperature control mode for RET”)
Set the start point for the heating function by pressing the rotary knob (C)
The set-point and actual temperatures will be shown alternately on the display (E)
When the heating is switched on, the LED “Heating plate” (G) and the LED “Set-point value” (I)
will be lit.
During agitation and standby operation, the display (E) will show if the base plate temperature
is above 50 °C.
HOt
The maximum temperature that can be set for the heating plate is 310 °C
i
i
5000800a_RET_REO_032008:6 23.06.2008 16:26 Uhr Seite 17
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com