Programs:
In the “Programs“ menu option, you can create 5 user-defined “Temperature / Time” profiles�
In addition, you can define the control mode (Time, Temperature hysteresis and Ramp) in the
program�
A program can consist of up to 5 segments�
Once a program has been selected, the following options are available:
Start:
Start the selected program upon request�
When you start the selected program, you are required to select "Infinite loop" or input "Loop
count"�
1� Infinite loop:
Upon completion of the last segment, the program continues with the first
segment until the user ends the program by stopping a device function�
2� Loop count:
Indicates the total number of loop cycles until program end�
Press “OK” to start the program control� Confirm the following message if you want to start
the program control, or press the “Menu” or “Back” button to end the process�
Note:
At the end of the program all device functions (tempering function) are switched off�
Edit:
Edit the selected program parameters�
Edit the selected program by pressing on the “Edit“ with tempering button (G)�
In the editing program, you can define up to 5 program segments� The selected program
segment is highlighted in blue color�
Then, you can edit, delete, insert or save the selected program segment in this program�
When editing a parameter, press the tempering button on the selected parameter� The
highlighted cursor changed into yellow color indicates that parameter can be edited�
If the “Back” button is pressed after editing, a reminding appears and ask if you want to save
the change� When you edit the program time for at least one segment, a symbol ( ) for
respective program will appear�
1� Seg No�: Segment number�
2� Temp�:
Target temperature�
3� Ctrl� Mode: Control Mode�
In control mode "Time" the target values and setting s of segment are valid for the duration
indicated in the column "Time"� Afterwards, the next program segment is automatically
executed�
Graph:
X - Axis:
In the "X - Axis" menu, you can set 5, 10, 20, 30, 60, 90 and 120 minutes as the scale for the
X - Axis�
Y - Axis:
In the "Y - Axis" menu, you can select "Automatic" or "Manual" axis scaling�
When you select "Automatic" axis scaling, a check mark (√) indicates that the function is activated�
In "Automatic" axis scaling, Y-axis is scaled to maximum value�
When you select "Manual" axis scaling, you need set the maximum and minimum value for
Y-Axis�
In control mode "± x�x K": The hysteresis (tolerance) of the actual temperature to target
temperature is set (e�g� ± 0�1 K)� The target values and settings of the segment are valid until
the actual temperature reaches the target tempe/- hysteresis for the first time�
Afterwards, the next program segment is automatically executed�
Control mode "RAMP" is allowed to moderate evenly with a defined slope� The segment is
closed after the target temperature has been reached�
The gradient in K/min comes from the quotient of the temperature difference and the time
specified in the current segment�
The temperature difference is calculated from the target value temperatures of the current and
preceding segment�
Note:
The ramp function can only be selected from the second segment�
Delete:
Deletes the selected program�
If a selected program is deleted by pressing on menu option “Delete” with the tempering button,
all the program parameters will be emptied� The symbol ( ) disappears�
Rename:
Rename the selected program�
You can rename the selected program by touching the tempering button on "Rename"�
Then, you can input a new name as required for the program with the keyboard appeared
subsequently�
Safety:
Password:
In the “Password” menu option, you can lock the menu settings by a 3-digit password (factory
setting: 000)�
If you forgot your password, please contact our service department�