9
ETSD5 0306
Page
Safety instructions
9
Correct use
10
Unpacking
10
Factory settings
10
Installation
10
Operating modes
11
Setting the HI-TEMP
11
Adjusting the set-point temperature
11
Replacing extension cable and sensor
11
Error messages
12
Accessories
12
Technical data
12
Caution!
The supply voltage (Limited Power Supply) for the ETS-
D5 must be provided using an IEC 61558 compliant transformer
with double or reinforced insulation. Alternatively, laboratory
equipment that meets the requirements of IEC 61010 or NEC
Class 2 may be used.
Do not touch the temperature sensor while measurements are
being taken. This will prevent incorrect results.
Always use the extension cable H 70 when the media being proces-
sed produces vapour. This ensures that the control unit does not
come into contact with the vapour.
Note:
This equipment must only be used with media that have a
flash point higher than the temperature safety limit set for the
magnet agitator and/or heating system used.
The equipment must not be operated in areas where there is a risk
of explosion.
Only use glass encapsulated temperature sensors for electrolysis
Contens
Safety instructions
EN
procedures.
The stainless steel temperature sensor must not be used with
aggressive media such as acids, caustic solutions or distilled
water, due to the risk of corrosion. The H 66 glass sensor should
be used in such cases.
Please ensure that the spiral cable does not come into contact
with the heating plate.
Take care when touching the temperature sensor!
Burn hazard!
The temperature sensor may heat up to 450 °C
when inserted in the media!
Please observe all accident prevention regulations applicable to
the activity and the work station. Personal protective equipment
should be worn when operating the equipment.
Danger:
The user must check that all conditions have been met for the safe
and correct operation of the equipment before operating the IKA-
TRON
®
ETS-D5 with a heater that does have a DIN 12878 compli-
ant connector, and therefore does not provide cut-out proection.
(The cut-out protection is a mechanism that ensures that the elec-
trical heating circuit is switched off should the contact thermome-
ter (ETS-D5) be broken or if the connector is removed.).
H+P magnetic agitator systems (no break fuse protection inclu-
ded)
When the supply current for the ETS-D5 is interrupted, the elec-
tronics for the heatable magnetic agitator system will take over
the temperature control. This means that the magnetic agitator
system will heat up to the temperature set (e.g. 300 °C). Readjust
the set-point temperature for the magnetic agitator system or
switch the heating off.
Danger when using flammable media with a low boiling
temperature.
Corning Magnetic Agitator PC-400, PC-420, PC-600, PC-620
When using the ETS-D5 in conjunction with a Corning Magnetic
Agitator it is absolutely essential that the rotary "heat" dial on the
magnetic agitator is set to the "O" (off) position. The ETS-D5 will
3564800b_ETSD5_032007:ETS-D4_a 18.06.2008 12:14 Uhr Seite 9
Содержание IKATRON ETS-D5
Страница 3: ...ETSD5 0306 Fig 1 Fig 2 3 Fig 3 Fig 4 Fig 5 3564800b_ETSD5_032007 ETS D4_a 18 06 2008 12 14 Uhr Seite 3...
Страница 17: ...17 ETSD5 0306 3564800b_ETSD5_032007 ETS D4_a 18 06 2008 12 14 Uhr Seite 17...
Страница 35: ...ETS D5 ETS D5 35 ETSD5 0306 3564800b_ETSD5_032007 ETS D4_a 18 06 2008 12 15 Uhr Seite 35...