The following error messages may be displayed (Fig. 5) and have
the causes listed below:
Er 1 Sensor not connected or defective (open circuit).
Er 5 Sensor not in media (no change in temperature of the
media after heater active for 3 minutes).
Er 6 Sensor short circuit.
Er 7 Maximum temperature for the media exceeded
(HI-TEMP setting).
Er 8 Calibration error (re-calibration by field service required).
H 16 V
Support rod
H 38
Retaining bar
H 44
Cross sleeve
H 50
Twin-jack adapter
H 52
Power supply kit
H 62
Standard sensor 270 mm stainless steel
H 66
Glass sensor
270 mm glass
H 70
Extension cable, 1 m long
AD-C1
Adapter (for Corning magnetic stirrer)
Measuring range:
°C
-50 ... 450
Resolution:
K
0.1)
Measuring accuracy:
K
±0,2 + sensor tolerance PT 1000
DIN IEC 751 Class A
Setting accuracy:
K
0,1
Type of control:
PID
Control deviation:
K
± 0,5
(500 ml water in 600 ml glass beaker,
magnetic stirring bar 40, 600 min-
1
, 50 °C)
Supply voltage:
V DC 8 ... 16 (Limited Power Supply)
Current consumption:
mA
8 (at 9V)
Ambient temperature:
°C
0 ... 60
Relative humidity:
%
80
Duty cycle:
%
100
Protection type:
IP 54
Cable length:
mm 350
Male connector:
6-pin DIN 45322
Socket:
DIN 12878 Class 2
Dimensions: (W x D x H) mm 80 x 20 x 90 (without sensor)
Weight:
kg
0.2
12
ETSD5 0306
Error messages
Accessories
Technical data
3564800b_ETSD5_032007:ETS-D4_a 18.06.2008 12:14 Uhr Seite 12
Содержание IKATRON ETS-D5
Страница 3: ...ETSD5 0306 Fig 1 Fig 2 3 Fig 3 Fig 4 Fig 5 3564800b_ETSD5_032007 ETS D4_a 18 06 2008 12 14 Uhr Seite 3...
Страница 17: ...17 ETSD5 0306 3564800b_ETSD5_032007 ETS D4_a 18 06 2008 12 14 Uhr Seite 17...
Страница 35: ...ETS D5 ETS D5 35 ETSD5 0306 3564800b_ETSD5_032007 ETS D4_a 18 06 2008 12 15 Uhr Seite 35...