13.6  Menu (Details):

13.6.1 Main menu (Explanation of functions)

Stirrer

Pressing the rotary/pressing knob (K) starts and stops the 

stirrer function. The speed can be set during operation.

Interval modus:

The menu allows the user to activate the „run/stop” and

 

„CW/CCW“ function. The operating and stop times can be 

separately configured.

The mode “run/stop” allows the stirrer to run clockwise 

during time 1 and stops the stirrer during time 2.

The mode “CW/CCW” allows the stirrer to run clockwise 

during time 1 and counterclockwise during time 2.

Speed limit:

The menu allows the user to set the desired maximum 

upper speed limit for the stirrer. The factory setting corre-

sponds to the maximum permitted speed of the stirrer. If 

the limit is changed, then the speed can only be selected 

within the set range.

Temperature

Temperature limit:

The factory setting corresponds to the maximum permitted 

temperature of the device. If the limit is changed, then the 

value can only be selected within the set range.

Calibration:

In chapter “14 Operation, Calibration” you can calibrate 

the external temperature sensor.

Reset calibration:

This option is used to reset the temperature measuring re-

ference value.

pH sensor

This menu allows the user to carry out settings for the 

pH control. The gas valves and pumps can supply the 

corresponding gases of liquids to the reactor container of 

the Algaemaster 10 in order to change the pH value.

Connect sensor:

 NOTICE

The pH sensor is not auto-

matically detected by the 

controller. Connection is 

carried out manually in the 

menu.

Calibration:

The pH sensor must be calibrated before starting a pH 

measurement.

In chapter “14 Operation, Calibration” you can calibrate 

the pH sensor.

Reset calibration:

This option is used to reset the pH measuring reference 

value.

Pumps

Pump limits:

By changing the limits, the set value of the flow for the 

pumps can be set within the limit range.

Calibration:

In chapter “14 Operation, Calibration” you can calibrate 

the pumps.

Reset Calibration:

This option is used to reset the pump measuring reference 

value.

Timer

Time

:

The set time indicates how long the timer should run.

Actions after expiry:

This setting defines which functions should be carried out 

after the time has expired.

Activate

:

Ticking the “Activate” box switches the timer on. If the func-

tion remains inactive, the counter is automatically active.

Start/Stop:

Ticking the box starts/stops the timer. The set time is dis-

played in the status bar of the working screen.

File manager

In this menu, you can manage the measuring files stored 

internally. For further information, see chapter “14 Opera-

tion, File manager”.

Safety

Password:

In the ˝Password“ menu, the menu settings can be locked 

by a 3-digit password (Factory settings: 000). 

Settings

Languages:

The “Languages“ option allows the user to select the de-

sired language by turning and pressing the rotary/pressing 

knob (K). A tick indicates the language that is set for the 

system.

Units:

Here you can specify the measuring units for the tempera-

ture values in °C or °F. A tick indicates the unit that is set 

for the system.

Display:

The ”Display“ option allows the user to change the back-

ground color and brightness of the working screen. 

Ticking the “Firmware Update Info” box, an info screen for 

the Firmware Update Tool is displayed when the device is 

started.

Sound:

The ”Sound” option allows the user to activate/deactivate 

the key-press sound and to set the volume.

Date and time:

This menu allows the user to carry out settings for the date 

and time.

 NOTICE

If the Algaemaster 10 is 

switched off, the details for 

the date and time are stored 

for a maximum of six 

weeks. 

Содержание Algaemaster 10

Страница 1: ...IKA Algaemaster 10 IKA Algaemaster 10_082017 20000017374 Operating instruction EN 6...

Страница 2: ...2...

Страница 3: ...ion 1 Motor 2 Vessel lid 3 LED panel 4 Reactor vessel 5 Base plate 6 Controller Fig 1 Pos Description 7 Control elements and display 8 USB port data export 9 Hose pump 1 10 Hose pump 1 11 Needle valve...

Страница 4: ...on 16 Connection sockets for LED panel 17 RS 232 interface thermostat communication 18 Connection socket for motor 19 Power supply connection with fuse and switch 20 Protective caps Pos Description 21...

Страница 5: ...5 2 Dangerous spots Fig 6...

Страница 6: ...1 2 3 LED panel 11 11 2 4 Gas hoses 12 11 2 5 Flow meter 12 11 2 6 Temperature sensor 12 11 2 7 pH sensor 13 11 2 8 Pump cassette 13 11 2 9 Motor 13 11 2 10 Main socket 13 11 2 11 Thermostat 13 12 Ope...

Страница 7: ...ms with the following standards or normative documents EN 61010 1 EN 61010 2 051 EN 60204 1 EN 61326 1 EN 60529 and EN ISO 12100 4 Warranty In accordance with IKA warranty conditions the warranty peri...

Страница 8: ...otective ground contact The voltage stated on the type plate must correspond to the mains voltage After an interruption to the power supply the device starts up again The functions previously set do n...

Страница 9: ...quipment parts also constitute a hazard CAUTION The stirrer must only be operated when the reactor vessel is fully closed You must ensure that the stirring element is securely clamped Gradually increa...

Страница 10: ...k motor coupling and base plate Motor 2 LED panels Main cable H 11 9 Unpacking 10 Motor protection The stirring instrument is suitable for continuous operation The motor current is limited electronica...

Страница 11: ...can be as signed freely as required 1 The 4 connection sockets for the LED panels Fig 3 16 can be assigned freely as required 2 Place the reactor container onto the base plate and install the LED pane...

Страница 12: ...s during shipping the float might get stuck in the tube In this case flush the flow meter with dry and clean gas The flow of gas 1 2 Fig 2 12 can be regulated using the needle valves Fig 2 11 Clockwis...

Страница 13: ...ose end of the output to a plug with grommet on the vessel lid Fig 5 26 11 2 9 Motor 1 Insert the motor into the corresponding holder for the motor coupling on the vessel lid Fig 5 25 2 Turn the motor...

Страница 14: ...For navigating through the menus selection and changing settings L Selection key pump 2 Start Stop pump 2 M Selection key for pump edit menu Open the edit menu for the pump functions selection of menu...

Страница 15: ...Observe the ambient conditions listed under 18 Technical Data After switching off via the main switch Fig 3 19 the device type device designation user defined device name and the firm ware version are...

Страница 16: ...for gas 1 and gas 2 d Gas valve 1 status Displays whether gas valve 1 is closed red or open green e Gas valve 2 status Displays whether gas valve 2 is closed red or open green f Pump 1 Displays whethe...

Страница 17: ...ect the menu by turning the rotating pressing knob K to the right or left to select the desired menu or sub menu which can then be selected by pressing the rotating pressing knob K Push or turn the ro...

Страница 18: ...rking screen If you press the Back key H the system skips back to the previous display Menu navigation Press the Menu key Fig 8 I and rotate rotating pressing knob Fig 8 K If you press the Back key Fi...

Страница 19: ...at can be stored Set Speed activated Actual Speed activated Actual direction of rotation stirrer activated Actual light intensity activated Set Temperature activated Actual thermostat temperature acti...

Страница 20: ...ommunication Device Name Algaemaster 10 Bluetooth deactivated Information Firmware Update Tool Display 1 Version Display 2 Version Logic Version Temperature C activated F deactivated Security Password...

Страница 21: ...it 7 0 Action above Hysteresis 0 05 Gas 2 pH Control deactivated Limit 7 0 Action above Hysteresis 0 05 Pumps Pump 1 Flow Speed 60 Activate Timer deactivated Time 00 00 00 00 dd hh mm ss Interval Mode...

Страница 22: ...libration This option is used to reset the pH measuring reference value Pumps Pump limits By changing the limits the set value of the flow for the pumps can be set within the limit range Calibration I...

Страница 23: ...nding function starts as soon as you press the Start button After time 1 has expired the function stops and time 2 begins to run with the corresponding function After time 2 has expired time 1 starts...

Страница 24: ...ime Select the selection button Q on the main screen Fig 8 Pos Menu Selection a Activate timer On Off b Time setting of timer Setting the time c Activate interval On Off d Time 1 Setting the time 1 in...

Страница 25: ...t Valve open pump starts Value above limit Valve close pump stops Value below limit Valve close pump stops Value above limit Valve open pump starts Hysteresis b Menu details pH Main menu pH see chapte...

Страница 26: ...g knob K to OK and confirm the entry by pressing 7 Immerse the pH sensor into the reactor container and deter mine the water temperature 8 Enter the measured pH value calibration value of the refe ren...

Страница 27: ...er from a filled vessel 1 into an empty vessel 2 for a period of one minute The pumped volume is referred to as a calibration value NOTICE Vessel 2 must at least correspond to the volume of the filled...

Страница 28: ...the current memory available Separator Delineation of multiple values within the csv file with comma or semicolon 1 Insert a USB memory stick with sufficient memory into the USB interface Fig 2 6 2 S...

Страница 29: ...lugging Connected devices and their properties are automatically recognized First download the latest driver for IKA devices with USB interface from http www ika com ika lws download usb driver zip an...

Страница 30: ...positi on isn t ok Check position of temperature probe Error 4 Stirrer is blocked No stirrer rotation Stirrer is blocked Unblock stirrer Error 6 Device Temperature Error Analysis of internal PCB temp...

Страница 31: ...A appliances using these IKA approved cleaning agents Dirt Cleaning agent Dyes Isopropanol Building materials Water containing detergent Isopropanol Cosmetics Water containing detergent Isopropanol Fo...

Страница 32: ...temperature sensor pH sensor min ml 6000 Cooling medium temperature min C 4 Cooling medium temperature max C 60 Cooling medium pressure max bar 0 5 Dimensions x H mm 260 x 450 Reactor lid Material PE...

Страница 33: ...13 pH measurement resolution pH 0 01 Connection for pH sensor BNC Interfaces data USB 1 0 2 0 type A socket data transfer USB stick Frontside USB 1 0 2 0 type B socket PC control Backside RS 232 PC c...

Страница 34: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com...

Отзывы: