1.4.4 Basic safety information during normal operation
Danger
Before switching on the machine, obtain information on
how to behave in the event of abnormal occurrences.
Danger
The machine may only be operated by appropriately
trained and authorised persons familiar with the operat-
ing instructions (and those for the subassemblies) and
capable of working according to them.
Danger
At working chamber temperatures exceeding 65°C, there
is a danger of burning at the container and supply pipes.
In this case, the operator must apply protection against
accidental contact with hot components.
Caution
The machine may not run dry!
The mechanical seal must always be surrounded by
liquid.
Info
The machine is not suitable for adding and processing
purely dry substances. This can result in serious dam-
age to the device.
Info
When changing the dispersion material, check whether
the materials of the axial-face seal and O-ring are still
suitable. It may be necessary to replace these compo-
nents with other materials. If in doubt, contact the sup-
plier.
Info
Following all assembly work, free movement of the drive
shaft must be ensured by manual turning before switch-
ing on the drive. Secure the drive against unauthorised
activation.
Содержание 06-5819
Страница 1: ...IKA WERKE Reg No 4343 01 IKA ULTRA TURRAX UTL 1000 10 Serial No 06 5819 OPERATING INSTRUCTIONS EN ...
Страница 2: ...IKA WERKE 2 ...
Страница 5: ...IKA WERKE 3 ...
Страница 36: ...IKA WERKE 34 ...
Страница 40: ...IKA WERKE 38 12 Assembly drawing ...
Страница 41: ...IKA WERKE 39 12 1Detail 1 of assembly drawing ...
Страница 42: ...IKA WERKE 40 12 2Detail 2 of assembly drawing only with option base frame ...
Страница 43: ...IKA WERKE 41 13 Dimension sheet ...
Страница 46: ...IKA WERKE 44 ...