15
Intended use
Use:
When used in combination with one of our recommended dispersing
elements, the drive unit is a high-speed dispersing and emulsifying unit
capable of handling free-flowing and liquid media in batches.
- Production of:
Emulsions
Dispersions
Wet
crushing
- Operating modes: Hand hold mode
On
stand
Range of use:
Indoor environments similar to that a laboratory of research,
teaching, trade or industry.
The safety of the user cannot be guaranteed:
- If the instrument is operated with accessories that are not supplied
or recommended by the manufacturer
- If the instrument is operated improperly or contrary to the manu-
facture’s specifications
- If the instrument or the printed circuit board are modified by the
third parties.
• Powder can not be too close to the flange. It can be blown away
by air turbulences of the drive.
• The instrument starts up again following a cut in the power sup-
ply. Please note that the instrument must be switched off at the
ON/OFF switch in this case before the power supply is recon-
nected.
• The instrument can only be disconnected from the mains supply
by pulling out the mains plug or the connector plug.
• The socket for the mains cord must be easily accessible.
• Always disconnect the plug before fitting accessories.
• Safe operation is only guaranteed with the accessories described
in the ”
Accessories
” chapter.
Never run dispersion tools dry, as the gasket
and bearings will be destroyed if the tools are
not cooled by the medium.
• The voltage stated on the nameplate must correspond to the
mains voltage.
• Protect the instrument and accessories from bumps and impacts.
• The instrument can only be opened by experts.
• Removable parts must be refitted to the instrument to prevent
the infiltration of foreign objects, liquids etc.
• Do not cover the ventilation slots on the drive in order to ensure
adequate cooling of the drive.
• The gasket and bearings are made of PTFE and rustproof steel; the
following points should therefore be noted:
Chemical reactions of
PTFE occur in contact with molten or solute alkali metals and alka-
line earth metals, as well as with fine powders of metals in groups
2 and 3 of the periodic system at temperatures above 300 °C - 400
°C. Only elementary fluorine, chlorotrifluoride and alkali metals at-
tack it; halogenated hydrocarbons have a reversible swelling effect.
(Source: Römpps Chemie-Lexikon and ”
Ulmann
”, Volume 19)
WARNING
Содержание 0003737000
Страница 3: ......
Страница 47: ...46 Fig 1 IKA 20 mm 10mm Fig 1 IKA...
Страница 48: ...47 PTFE 300 400 C PTFE PTFE Source R mpps Chemie Lexikon and Ulmann Volume 19 T 10 R 200 6 20m s...
Страница 53: ...52 JA 52 53 54 55 55 55 55 56 56 56 57 57 58 58 59...
Страница 54: ...53 1 IKA 20mm 10mm...
Страница 55: ...54 1 ON OFF IKA PTFE PTFE 300 400 R mpps Chemie Lexikon and Ulmann Volume 19...
Страница 57: ...56 1 3 1 3 1 3 10mm T10 1 2 1 1 1 1 T10 8 000 30 000RPM ON OFF 1 2 T10 ON OFF T10...
Страница 89: ...88 o o 1 IKA 20 mm 10 mm Fig 1 BG...
Страница 90: ...89 IKA PTFE 300 C 400 C PTFE 2 3 19...
Страница 93: ...92 o o Fig 1 IKA 20 mm 10 mm Fig 1 EL...
Страница 94: ...93 IKA PTFE PTFE 2 3 300 C 400 C R mpps Chemie Lexikon Ulmann 19...
Страница 95: ...94 Note...
Страница 96: ......