Iijima MA-300 Скачать руководство пользователя страница 21

 

19 

MA300TS-0210 

8. Error Messages 

 

 

 

 

 

 

    

 

An error message may display during the self-diagnosis during use. Check the details of the displayed error, and 
troubleshoot with the following actions. 

Symptom 

Confirmation 

Action 

Sensor error 

 
 
 

 
 
 

 

  The sensor output is unstable 

during span calibration or zero 

calibration. 

  The sensor output is unstable 

during measurement. 

  The ambient temperature has 

changed drastically. 

  The oxygen sensor 

membrane is damaged. 

Is the oxygen sensor correctly 

mounted? 

Securely connect the oxygen sensor 

(WA-BRM5) to “DO IN” on the back panel of 
the PACK MASTER. (Refer to page 5.) 

Has the temperature of the 
measurement environment 

changed drastically? (Temperature 
difference between storage site 
and measurement site, 

temperature change caused by air 
conditioning, etc.) 

The device cannot be calibrated correctly if 
the temperature changes drastically. 

Calibrate the span after allowing the 
temperature to stabilize sufficiently in a stable 
ambient temperature, such as before turning 

on the air conditioning. 

Was the membrane on the tip of 

the oxygen sensor facing upward 
when calibrating the span? 

Point the membrane on the tip of the oxygen 

sensor downward or to the side when 
calibrating the span. 

Was stabilizing time taken after 

replacing the oxygen sensor 
membrane and electrolytic 
solution? 

When the membrane and electrolytic solution 

have been replaced, the membrane tension 
is not stable, so allow the membrane to settle 
for several hours to 12 hours. Calibrate after 

this stabilizing time. 

Is the oxygen sensor’s membrane 
surface damaged or wrinkled, 

etc.? 
 

Refer to the “Maintenance Guide” enclosed 
with the oxygen sensor and replace the 

membrane and electrolytic solution with new 
ones. After replacing, allow the parts to 
stabilize for several hours to 12 hours or 

more, and then calibrate. 

Is the membrane on the tip of the 

oxygen sensor contaminated? 

Clean the oxygen sensor and remove the 

contaminants. (Refer to section “6. 
Maintenance (1) Cleaning after 
measurement” on page 16.) 

If the contaminants cannot be removed, 
replace the membrane and electrolytic 
solution with new ones. After replacing, allow 

the parts to stabilize for several hours to 12 
hours or more, and then calibrate. 

Is the oxygen sensor being used 
after being suspended for a long 

time (several months)? 

The oxygen sensor membrane and 
electrolytic solution may be degraded. 

Replace the membrane and electrolytic 
solution with new ones. After replacing, allow 
the sensor to stabilize for several hours to 12 

hours or more, and then calibrate it. 

If the symptoms do not improve 

after taking the above actions, the 
oxygen sensor may have reached 
its life. 

Order a new oxygen sensor and replace the 

old one. 

If the symptoms do not improve 

even after replacing the oxygen 
sensor, the main unit may be 
faulty. 

Make a request for repairs. 

 
If any other error message is displayed, refer to “Error Messages” section in the PACK MASTER Operation Manual. 

Содержание MA-300

Страница 1: ... DO in fluid using the optional residual oxygen meter PACK MASTER model RO 105S and a stirrer This manual describes how to operate the DO Measuring Device and the precautions that must be observed to ensure proper and safe use Always read this manual and fully understand the operation before starting use Keep this manual near the device for quick reference at any time ...

Страница 2: ...pan Calibration Method Saturated water calibration air calibration 8 2 Span calibration 9 1 Air calibration method 9 2 Saturated water calibration method 10 3 Zero calibration 11 4 Water temperature adjustment 12 5 Measurement 13 1 Measuring a sample Sample that does not contain fibers or solids 13 2 Repeated measurement 15 3 Measuring a sample containing fibers or solids 15 6 Maintenance 16 1 Cle...

Страница 3: ...urement will not be possible Exchange the 3 way check valve with the enclosed 3 way cock Refer to section 5 Measurement 3 Measuring a sample containing fibers or solids on page 15 of this manual The key components are made of acrylic materials Acrylic materials have a low resistance to solvents such as alcohol so do not measure such matters The acrylic material could be adversely affected such as ...

Страница 4: ...KOKUGO Co Ltd Part No 02 048 01 02 3 way check valve x 1 pc Manufacturer ISIS Co Ltd Part No VCV1251 1 set 4 3 way cock Part for measuring samples containing fibers or solids Model VXB1079 Manufacturer ISIS Co Ltd Material Polycarbonate polyethylene 1 5 Cleaning brush Part name Blood sedimentation brush Model 7 5611 01 Distributor AS ONE Corporation 1 6 Anhydrous sodium sulfite Guaranteed Reagent ...

Страница 5: ...3 MA300TS 0210 2 Accessories options Description 8 Stirrer model KK Product name Ultra Flat Digital Mini Magnetic Stirrer Distributor KENIS Ltd ...

Страница 6: ...Do not use a tool when connecting the parts Tightening with a tool could result in part damage Securely tighten the parts by hand without using tools Schematic drawing PACK MASTER and stirrer are sold separately The following photo shows the overall view including the optional PACK MASTER model RO 105S Stirrer Model KK and sample prepared by the user DO sensor connector DO sensor connector The mod...

Страница 7: ...MASTER Remove the pin shorting cap from the oxygen sensor and connect the connector part Carefully align the connector pins with the position of the hole Connect the connector and turn the coupling to fix it Step 2 Connect the tubing between the syringe and housing Step 3 Install the stirrer Place the magnetic stir bar in the housing Place the housing over the stirrer and turn the stirrer power ON...

Страница 8: ...tting Step 1 Confirm that the oxygen sensor WA BRM5 is connected Step 2 Turn the PACK MASTER power ON and press on the Measurement Standby screen The mode switches to the DO MEASURE mode When O2 MEASURE is displayed When DO MEASURE is displayed The display alternates in the following order each time is pressed DO MEASURE cannot be selected unless the oxygen sensor WA BRM5 is connected to the PACK ...

Страница 9: ...3 Press The SETTING MENU screen opens and O2 DISP SET is selected flashes Step 4 Press three times The cursor moves and DO DISP SET is selected flashes Step 5 Press The DO DISP SET screen opens and the selected setting flashes The factory default setting is DO mg L Step 6 Each time is pressed the display changes as shown below When DO mg L is displayed When SATUR R is displayed When set to DO mg L...

Страница 10: ...en opens and O2 DISP SET is selected flashes Step 4 Press four times The cursor moves and DO CAL SET is selected flashes Step 5 Press The DO CAL SET screen opens and the selected setting flashes The factory default setting is AIR Step 6 Each time is pressed the display changes between AIR and SATUR WATER When AIR is displayed When SATUR WATER is displayed Display symbol Name AIR Air calibration SA...

Страница 11: ...stable such as before turning on the air conditioning Step 2 Change the instrument setting to Set to AIR Perform only the first time Refer to section 3 Setting the Span Calibration Method Saturated water calibration air calibration on page 8 of this manual Step 3 When the DO MEASURE screen is displayed press The MENU screen opens and CAL is selected flashes Step 4 Press The DO CAL screen opens Ste...

Страница 12: ... Allow five to ten minutes for the temperature to stabilize If the temperature varies it can take approximately one minute per degree Celsius to stabilize Step 5 When the DO MEASURE screen is displayed press The MENU screen opens and CAL is selected flashes Step 6 Press The DO CAL screen opens Step 7 When the indicated value stabilizes hold down for 1 second or longer The stability is decided and ...

Страница 13: ...ected flashes Step 4 Press The DO CAL screen opens Step 5 When the indicated value stabilizes hold down for 1 second or longer The stability is decided and after approx 10 seconds CAL complete appears to indicate that the calibration is complete The DO MEASURE screen opens when the automatic calibration is completed The calibration value is saved even if the PACK MASTER power is turned OFF the bat...

Страница 14: ...O2 Measurement Standby screen or DO MEASURE screen The MENU screen opens and CAL is selected flashes Step 3 Press twice The cursor moves and FUNC is selected flashes Step 4 Press The FUNC MENU screen opens and CLOG CHECK is selected flashes Step 5 Press four times The cursor moves and WTR TEMPADJ is selected flashes Step 6 Press The DO C1 P ADJUST screen opens Step 7 Change the water temperature t...

Страница 15: ...rs or solids refer to section 5 Measurement 3 Measuring a sample containing fibers or solids on page 15 of this manual 1 Measuring a sample Sample that does not contain fibers or solids Step 1 Carefully mount the oxygen sensor WA BRM5 on the housing unit Do not insert the oxygen sensor forcefully Forcefully inserting the sensor can apply a load on the membrane and affect the measurement value If t...

Страница 16: ...d or canceled If the stirrer is rotated at high speed the magnetic stir bar may pop off with force and damage the oxygen sensor Gradually increase the stirrer speed If the stirrer speed is too slow the flow will be insufficient and the measurement value may be low Adjust the speed to the maximum speed that does not cause the magnetic stir bar to pop off The recommended speed for the stirrer Model ...

Страница 17: ... atmosphere will cause the measurement value to be higher 3 Measuring a sample containing fibers or solids Step 1 Exchange the 3 way check valve with the 3 way cock 3 way check valve 3 way cock Step 2 Take the measurement while switching the lever position for suctioning and discharge Refer to the syringe operation explained in Step 3 of section 5 Measurement 1 Measuring a sample on page 14 of thi...

Страница 18: ...s a beaker with tap water and lightly rinse the oxygen sensor in the container Always clean the oxygen sensor with tap water filled in a container If cleaned with tap water directly from the faucet the pressure of the water could break the membrane on the sensor surface Step 2 Wipe off the moisture lightly with tissue paper and store the sensor in the atmosphere When removing the oxygen sensor fro...

Страница 19: ...he membrane on the top of the oxygen sensor contaminated Clean the oxygen sensor and remove the contaminants Refer to section 6 Maintenance 1 Cleaning after measurement on page 16 If the contaminants cannot be removed replace the membrane and electrolytic solution with new ones After replacing allow the sensor to stabilize for several hours to 12 hours or more and then calibrate it Was zero calibr...

Страница 20: ... have reached its life Order a new oxygen sensor and replace the old one Contaminants have solidified in the housing Fill the housing with lukewarm water and allow it to soak Then wipe off the contaminants with a soft sponge etc If the solution contains oils and fats soak the sensor in a neutral detergent and wipe off the contaminants with a soft sponge etc Do not use alcohol when cleaning Do not ...

Страница 21: ...been replaced the membrane tension is not stable so allow the membrane to settle for several hours to 12 hours Calibrate after this stabilizing time Is the oxygen sensor s membrane surface damaged or wrinkled etc Refer to the Maintenance Guide enclosed with the oxygen sensor and replace the membrane and electrolytic solution with new ones After replacing allow the parts to stabilize for several ho...

Страница 22: ...m which you purchased the product if you have any questions or inquiries regarding handling repairs or inspections 1 1 Ishida Toyooka cho Gamagori shi Aichi 443 0011 JAPAN Telephone 81 533 67 2827 Facsimile 81 533 69 6814 https www iijima e co jp ...

Отзывы: