3. Setting the Span Calibration Method (Saturated water calibration/air calibration)
Step 1. Turn the PACK MASTER power ON, and press on the O
2
Measurement Standby screen or DO
MEASURE screen.
The MENU screen opens, and “CAL” is selected (flashes).
Step 2. Press once.
The cursor moves, and “SET” is selected (flashes).
Step 3. Press .
The SETTING MENU screen opens, and “O
2
DISP SET” is selected
(flashes).
Step 4. Press four times.
The cursor moves, and “DO CAL SET” is selected (flashes).
Step 5. Press .
The DO CAL SET screen opens, and the selected setting flashes.
* The factory default setting is “AIR”.
Step 6. Each time is pressed, the display changes between “AIR” and “SATUR WATER”.
<When AIR is displayed>
<When SATUR WATER is displayed>
Display symbol
Name
AIR
Air calibration
SATUR WATER Saturated water
calibration
<When Set to AIR is selected>
Step 7. When is pressed, the selected settings are saved.
The SETTING MENU screen opens.