5. If an Irregular Heartbeat (IHB) is detected during the measurement procedure, the IHB symbol will be displayed in

the “iHealth MyVitals” APP. Under this condition, the Wireless Blood Pressure Monitor can keep functioning, but the

results may be inaccurate. Please consult your physician for accurate assessment.

The IHB symbol will be displayed under 2 conditions:

1) The coefficient of variation (CV) of pulse period >25%.

2) The difference of adjacent pulse period is ≥0.14s and more than 53 percent of the total number of pulses readings

falls within this definition.

6. Please do not use any cuff other than that supplied by the manufacturer as this may result in inaccurate

measurements.

7. For information regarding potential electromagnetic or other interference between the blood pressure monitor and

other devices together with advice regarding avoidance of such interference, please see ELECTROMAGNETIC

COMPATIBILITY INFORMATION. It is suggested that the blood pressure monitor be kept 10 meters away from other

wireless devices, such as WLAN unit, cell phone, microwave oven, etc.

8.This product should not be used as a USB device.

9. If the blood pressure measurement (systolic or diastolic)is outside the rated range specified in part

SPECIFICATIONS, the monitor will immediately display a technical alarm on the LED screen. In this case, repeat the

measurement ensuring that the proper measurement procedures are followed and/or consult with your medical

professional. The technical alarm is preset in the factory and cannot be adjusted or inactivated. This technical alarm is

assigned as low priority according to IEC 60601-1-8. The technical alarm does not need to be reset.

10. This device requires a medical AC adapter with an output of DC 5.0V that complies with IEC60601-1/UL 60601-1

and IEC 60601-1-2/EN 60601-1-2 such as UE05WCP-050100SPC(input:100-240V, 50/60Hz, 0.18A; output: DC 5V,

1.0A). Please note that the monitor jack size is USB microAB. The USB jack should be used for charging only.

11.Use of Charging Cable other than those specified or provided by the manufacturer of this equipment could result in

increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper

operation.

12.Measurements are not possible in patients with a high frequency of arrhythmias.

13.The device is not intended for use on neonates, children or pregnant women. (Clinical testing has not been

conducted on neonates, children or pregnant women.)

14.Motion, trembling, shivering may affect the measurement reading.

15.The device would not apply to the patients with poor peripheral circulation, noticeably low blood pressure, or low

body temperature (there will be low blood flow to the measurement position).

16.The device would not apply to the patients who use an artificial heart and lung (there will be no pulse)

17.Consult your physician before using the device for any of the following conditions: common arrhythmias such as

atrial or ventricular premature beats or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, pre-eclampsia, enal

diseases.

18.The patient is an intended operator.

This Monitor is designed for adults and should never be used on infants, young children, pregnant or

pre-eclamptic patients. Consult your physician before use on children.

This product might not meet its performance specifications if stored or used outside the specified temperature and

humidity ranges.

Please do not share the cuff with any infectious person to avoid cross-infection.

Please do not attempt to open or disassemble the product, under inappropriate operation, the device and battery can 

be damaged and hazardous to the human body and the environment.

Содержание Neo BP5S

Страница 1: ...DICATORS 2 SET UP REQUIREMENTS 3 SET UP PROCEDURES 3 MEASUREMENT PROCEDURES 3 TAKING YOUR BLOOD PRESSURE READING 5 SPECIFICATIONS 6 GENERAL SAFETY AND PRECAUTIONS 6 BATTERY HANDLING AND USAGE 8 TROUBL...

Страница 2: ...sure measurement system It is designed to measure the systolic and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around th...

Страница 3: ...e iHealth Cloud Account Your iHealth account also gives you access to the free and secure iHealth cloud service Go to www ihealthlabs com and sign in with the same account Power on the Monitor Press S...

Страница 4: ...s measured position the monitor in the middle of your arm so that it is aligned with your middle finger When the right arm is measured position the monitor in the middle of your arm so that it is alig...

Страница 5: ...Note Please consult a health care professional for interpretation of pressure measurements BLUETOOTH FUNCTION Connect to iOS Device via Bluetooth a Enable Bluetooth on your iOS device b Launch the iH...

Страница 6: ...r storage and transport 90 RH 17 Environmental pressure 80kPa 105kPa 18 Battery life more than 130 measurements on a full charge 19 The blood pressure measurement system includes accessories pump valv...

Страница 7: ...that complies with IEC60601 1 UL 60601 1 and IEC 60601 1 2 EN 60601 1 2 such as UE05WCP 050100SPC input 100 240V 50 60Hz 0 18A output DC 5V 1 0A Please note that the monitor jack size is USB microAB...

Страница 8: ...tery low charge symbol flashing for a few seconds Do not change the battery If the battery can no longer be charged please contact Customer Service Overcharging the battery may reduce its lifetime Lit...

Страница 9: ...Bluetooth communicate error Reset monitor by pressing the START STOP button and holding for about 10 seconds then connect the mobile device correctly and try again If the monitor is still abnormal ple...

Страница 10: ...ed with Ethyl alcohol 75 90 Then air dry the cuff 8 It is recommended that product performance be checked every 2 years or after each repair Please contact the customer service Do not drop this monito...

Страница 11: ...e equipment Note This product has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable prot...

Страница 12: ...tre antenne ou metteur CE Statement The CE Mark applies to products regulated by certain European health safety and environmental protection legislation The CE Mark is obligatory for products it appli...

Страница 13: ...ent Harmonic distortion IEC 61000 3 2 Class A Home healthcare environment Voltage fluctuations and flicker IEC 61000 3 3 Compliance Home healthcare environment Table 2 Enclosure Port Phenomenon Basic...

Страница 14: ...nts burst IEC 61000 4 4 2 kV 100kHz repetition frequency Surges Line to line IEC 61000 4 5 0 5 kV 1 kV Surges Line to ground IEC 61000 4 5 0 5 kV 1 kV 2 kV Conducted disturbances induced by RF fields...