D E V I C E T U N I N G
14
MRBP.424321.001G2 Version 1.0 15/06/2020
4.2
COMMUNICATION SETTINGS
To select the communication settings, it is necessary to enter
«
MENU
»
→
«
SETTINGS
» →
«
COMMUNICATION
».
COMMUNICATION
NETWORK ADDRESS: 1
RATE, baud 19200
SAVE
In the
«
COMMUNICATION
»
menu it is possible to set the network address and baud rate for
the RS-485 interface.
The appearance of an asterisk in the upper line of the display - "COMMUNICATION*" means
that the values of the network address and /or rate have been changed. To apply and save the set
values, it is necessary to select "SAVE". The asterisk on the top line disappears after saving the com-
munication settings.
4.2.1
«
NETWORK ADDRESS
»
In the
«
NETWORK ADDRESS
»
menu it is possible to set the network address value 1 to 247.
Edited category is chosen by using the buttons
and
.
The parameter is changed by using the buttons
and
.
4.2.2
«
RATE
»
RATE
1200
2400
4800
9600
19200
34800
57600
115200
In the
«
RATE
»