Ultrasonic flow meter
SU2020 SU8020 SU2021 SU8021 SU2621 SU8621
53
12 Troubleshooting
The unit has many self-diagnostic options. It monitors itself automatically during operation.
Warnings and error states are displayed even if the display is switched off. Error indications are also
available via IO-Link.
The status signals are classified according to NAMUR recommendation NE107.
If several diagnostic events occur simultaneously, only the diagnostic message of the event with the
highest priority is displayed.
In addition, warnings and error messages are displayed by the unit as follows:
•
switching signal when using OUT1 or OUT2 as diagnostic output.
See also
•
colour signal of the operating status LED.
See also:
If the measured temperature value fails, the process value for flow rate is still available.
Additional diagnostic functions are available via IO-Link
Ò
IO-Link interface description at
12.1 Warning messages
Display indication
LED display
Problem/remedy
• Display off
• Operating status LED: off
• Switching status LEDs: off
• Supply voltage too low.
u
Check the supply voltage.
• Display switched off.
u
Check whether [diS.B] = OFF and
change setting if necessary.
• Operating status LED switched off.
u
Check whether [LED.M] = OFF is set
and change setting if necessary.
• Display off
• Operating status LED: flashes red
Internal unit temperature too high.
u
Allow unit to cool down.
• Title line:
Short circuit OUT1/
OUT2
• Process value line: ---
• Operating status LED: flashes red
• Switching status LEDs: flash yellow
rapidly
Short circuit in both outputs.
u
Check OUT1 and OUT2 for short
circuit or excessive current.
• Title line:
Short circuit OUT1
• Process value line: ---
• Operating status LED: flashes red
• Switching status LED1: flashes yellow
rapidly
Short circuit output 1.
u
Check OUT1 for short circuit or
excessive current.
• Title line:
Short circuit OUT2
• Process value line: ---
• Operating status LED: flashes red
• Switching status LED2: flashes yellow
rapidly
Short circuit output 2.
u
Check OUT1 for short circuit or
excessive current.
• Title line:
Under limit
• Process value line:
UL
• Operating status LED: flashes red
Below the display range.
u
Check the measuring range.
• Title line:
Over limit
• Process value line:
OL
• Operating status LED: flashes red
Above the display range.
u
Check the measuring range.
• Title line:
Override active
• Process value line: ---
• Operating status LED: flashes red
A process value differs from the meas-
ured value. PV is set to “0” while over-
ride bit is set in PDOut.
u
Deactivate PDOut override.
• Title line:
Signal quality low
• Process value line:
LOW
• Operating status LED: lights blue
Signal quality low.
u
Remove unit and check for deposits.
u
Check application for interference (air
bubbles/particles).