SU2020 SU8020 SU2021 SU8021 SU2621 SU8621
Ultrasonic flow meter
24
10 Parameter setting
Parameter setting can be carried out via the IO-Link interface or via the operating elements on the
unit.
Parameters can be set before installation or during operation.
If you change parameters during operation, this will influence the function of the plant.
u
Ensure that there will be no malfunctions in your plant.
During parameter setting the unit remains in the operating mode. It continues to monitor with the
existing parameter until the parameter setting has been completed.
Depending on the parameter setting, the parameters available in the menu may change.
10.1 Parameter setting via the unit keys
CAUTION
If the medium temperature is above 50 °C (122 °F), parts of the housing can increase in
temperature to over 65 °C (149 °F).
w
Risk of burns
u
Do not touch the device with your hands.
u
If necessary, use a blunt object to make settings on the device.
Intention
Action
Change from the process value display to the main menu
[●]
Change to the submenu
Use [▼] to navigate to the sub-
menu, e.g. [EF], then [●]
Select the requested parameter
[▲] or [▼]
Change to the setting mode
[●]
Modify the parameter value
[▲] or [▼]
Apply the set parameter
[●]
Exit parameter setting without saving
[▲] + [▼]
Return to the next higher menu level
(repeat several times to reach process value display)
[▲] + [▼]
- or -
Use [▼] to navigate to [Back],
then [●]
Return to the process value display
> 30 seconds (timeout)
When changing from the process value display to the main menu or after resetting the unit, the
guided installation option is automatically displayed. You are asked whether you want a guided
installation.
Guided installation via an installation wizard (
.
10.2 Parameter setting via IO-Link
Requirements for parameter setting via the IO-Link interface:
ü
A suitable parameter setting software, e.g. ifm moneo|configure
ü
The Input Output Device Description (IODD) for the device, see
ü
One IO-Link master