1  Preliminary note

1.1  Key to the symbols

Instruction

>

Reaction or result

[…] Designation of keys, buttons or indications

Cross-reference
Important note 

Non-compliance may result in malfunction or interference�
Information\Supplementary note

2  Safety instructions

•  Read this document before setting up the product and keep it during the entire 

service life�

•  The product must be suitable for the corresponding applications and 

environmental conditions without any restrictions�

• 

Only use the product for its intended purpose (→ Functions and features).

• 

Only use the product for permissible media (→ Technical data). 

•  If the operating instructions or the technical data are not adhered to, personal 

injury and/or damage to property may occur� 

•  The manufacturer assumes no liability or warranty for any consequences 

caused by tampering with the product or incorrect use by the operator�

•  Installation, electrical connection, set-up, operation and maintenance of the 

product must be carried out by qualified personnel authorised by the machine 

operator�

•  The unit complies with the standard EN 61000-6-4 and is a class A product� 

The unit may cause radio interference in domestic areas� If interference occurs, 

the user must take appropriate actions�

•  Protect units and cables against damage�

Содержание LDL100

Страница 1: ...Operating instructions Conductivity sensor hygienic G1 2 LDL100 80277836 00 06 2019 UK...

Страница 2: ...Installation procedure of the sensor 10 6 3 Notes on the use in accordance with EHEDG 11 6 4 Notes on the use according to 3 A 12 7 Electrical connection 13 7 1 For the cULus scope of validity 13 8 Pa...

Страница 3: ...8 9 1 Function check 18 9 2 Operating and diagnostic messages via IO Link 19 9 3 Output response in different operating states 19 10 Technical data and scale drawing 19 11 Maintenance Transport 19 12...

Страница 4: ...nd features Only use the product for permissible media Technical data If the operating instructions or the technical data are not adhered to personal injury and or damage to property may occur The man...

Страница 5: ...nit detects the conductivity and the temperature of liquids in pipes and tank systems The unit is designed for direct contact with the medium A PC with USB IO Link master or a programmed memory plug o...

Страница 6: ...ia prone to formation of deposit Do not expose the probe tip to direct sun radiation UV radiation 5 Function 5 1 Measuring principle The unit operates on the conductive measuring principle It measures...

Страница 7: ...analogue output OUT2 can be configured 8 4 Curve of the analogue signal factory setting I mA L T 20 4 Curve of the analogue signal measuring range scaled I mA L T 20 4 ASP2 AEP2 L T Conductivity Temp...

Страница 8: ...ary for the configuration of the unit detailed information about process data structure diagnostic information parameter addresses and the necessary information about the required IO Link hardware and...

Страница 9: ...ly before or in rising pipes Provide for inlet and outlet pipe lengths 5 x DN 1 1 5 x DN 5 x DN S sources of interference DN pipe diameter 1 sensor Disturbances caused by bends valves or pipe reductio...

Страница 10: ...e installation steps differs Observe the notes in the installation instructions of the respective adapter 6 2 2 Installation procedure of the sensor Slide the supplied seal black O ring 1 Fig 6 1 over...

Страница 11: ...e data sheet Make sure that the unit is integrated into the system according to EHEDG Use self draining installation Only use process adapters permitted according to EHEDG with special seals required...

Страница 12: ...ocess connection must be provided with a leakage port When adapters with 3 A certification are used this is guaranteed The leakage port must be clearly visible and must be installed facing downwards f...

Страница 13: ...y must only be made via SELV PELV circuits A class 2 power supply can also be used and is not excluded The device shall be supplied from Limited Energy circuit source according cl 9 4 of UL 61010 1 3r...

Страница 14: ...he unit If necessary read sensor again Remove USB IO Link master and put the unit into operation 9 8 2 Parameter setting via the memory plug Via a memory plug Accessories a parameter set can be writte...

Страница 15: ...C reference temperature for measuring the conductivity The standard temperature can be adjusted by the user if necessary Setting range 15 35 C T Cmp Temperature compensation The conductivity is determ...

Страница 16: ...d point conductivity Setting range 100 15000 S cm hysteresis for ASP2 COND 50 for ASP2 COND Lo T Minimum value memory for the temperature Hi T Maximum value memory for the temperature Reset Hi T and L...

Страница 17: ...n parameter T Cmp to a medium with the temperature coefficient 3 0 K Example T Cmp 3 0 Make all other settings Transfer the sensor data to the unit 8 5 Temperature influence and temperature coefficien...

Страница 18: ...n Example of values taken down medium at 25 C 500 S cm medium at 45 C 800 S cm Temperature change 20 K 4 Calculate the change of the conductivity in percent The conductivity has increased by 300 S cm...

Страница 19: ...ical data sheet and scale drawing at www ifm com 11 Maintenance Transport Prevent any formation of deposits and soiling on the sensor element If cleaning it manually do not use any hard or sharp objec...

Страница 20: ...setting rEF T 25 C T Cmp 2 uni T C CGA 100 ou2 I SEL2 COND conductivity ASP2 TEMP 0 C AEP2 TEMP 150 C Offset TEMP 0 K ASP2 COND 0 S cm AEP2 COND 15000 S cm Lo T Hi T Lo C Hi C FOU2 OFF dAP 1 s S Tim...

Страница 21: ...21 UK More information at www ifm com...

Отзывы: