8.1.2 Full adjustment
►
Fill the tank until the level is at least 20 mm above the set detection range�
►
Click on [Adjust full tank]
�
> If no IO-Link error occurs, the adjustment was successful�
> Event [Parameter missing] is not indicated any more�
> All LEDs in the detection range are on�
►
Empty the tank to check whether the unit operates correctly�
8.1.3 Dynamic adjustment
►
Fill the tank until the level is in the measuring range�
►
Click on [Start dynamic tank adjustment]
�
> The running light of the LED bar graph goes out� The level detected is
indicated�
►
Check whether the level is detected correctly�During the dynamic adjustment,
the level may change within the valid range�
►
Click on [Stop dynamic tank adjustment]
> If no IO-Link error occurs, the adjustment was successful�
> Event [Parameter missing] is not indicated any more�
►
Empty the tank to check whether the unit operates correctly�
If the sensor elements have been limited, the valid measuring range for the
dynamic adjustment is changed accordingly�
The dynamic adjustment will not stop automatically� It must be stopped or
aborted via a command
�
8.1.4 Empty and full adjustment
►
Empty the tank until the level is at least 20 mm below the set detection range�
►
Click on [Adjust empty tank]
> If no IO-Link error occurs, the adjustment was successful�
> Event [Parameter missing] is not indicated any more�
> The running light of the LED bar graph goes out�
►
Fill the tank until the level is at least 20 mm above the set detection range�