The unit can be set to different standard conditions:
•
Via the menu item [rEF�P] the standard pressure is set, which serves as a refer-
ence for the measured and display values for volumetric flow (→ 9.5.8).
•
Via the menu item [rEF�T] the standard temperature is set, which serves as a
reference for the measured and display values for volumetric flow (→ 9.5.9).
4.9 Low flow cut-off (LFC)
With this function small volumetric flow quantities can be ignored (→ 9.5.5). Flows
below the LFC value have no effect on the display and output signals�
4.10 Customer-specific calibration (CGA)
The customer-specific calibration allows changing the gradient of the curve of
measured values� It influences the display and the outputs�
140 % MEW
100 % MEW
60 % MEW
MEW
Q
A
V1
V0
V2
A
Operating value for display
and output signals
Q
Volumetric flow
MEW Final value of the measuring
range
V0
Curve of measured values
with factory setting
V1,
V2
Curve of measured values
after calibration
The change in the gradient is indicated in per cent� Factory setting = 100 %� After
a change the calibration can be reset to factory setting (→ 9.5.7).
5 Installation
The rules and regulations for the installation and operation of compressed
air equipment must be observed�
Installation in pipes:
►
Insert the unit into a G1 process adapter and tighten the nut� Tightening torque
max� 50 Nm� Ensure that the unit is correctly oriented�
5.1 Installation location
•
Downstream of the cold dryer / near the load�