SRL Series

SAFETY PRECAUTIONS

The following general safety precautions must be observed during all phases of operation, service, 
and repair of this instrument. Failure to comply with these precautions or with specifi c WARNINGS 
elsewhere in this manual may impair the protection provided by the equipment. Such noncompliance 
would also violate safety standards of design, manufacture, and intended use of the instrument. 

IET Labs assumes no liability for the customer’s failure to comply with these precautions.

The SRL is an indoor use product.

DANGEROUS PROCEDURE WARNINGS

Comply with all WARNINGS - Procedures throughout in this manual and instructions on the 
instrument prevent you from potential hazard. These instructions contained in the warnings must 
be followed.

BEFORE APPLYING POWER
Verify that all safety precautions are taken. Make all connections to the instrument before applying 
power. Note the instrument’s external markings described under “Safety Symbols”.

     • DO NOT Operate in an Explosive Atmosphere

     • Do not operate the instrument in the presence of infl ammable gasses or fumes 
     • Operation of any electrical instrument in such an environment clearly constitutes a safety  
hazard
     • Use Caution around live circuits and whenever hazardous voltages > 45 V are present
     • Operators must not remove instrument covers 
     • Component replacement and internal adjustments must be made by qualifi ed maintenance 
personnel only 
     • DO NOT substitute parts or modify the instrument
      • When working with high voltages; post warning signs, train personnel and keep unauthorized 
personnel away.

To  avoid  the  danger  of  introducing  additional  hazards,  do  not  install  substitute  parts  or  perform 
unauthorized modifi cations to the instrument.  

Return the instrument to an IET Labs for service and repair to ensure that safety features are maintained 
in operational condition.

Содержание SRL Series

Страница 1: ...ASUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC SRL Series Resistance Standard Operation Manual Copyright 2022 IET Labs Inc Visit www ietlabs com for ma...

Страница 2: ...PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC...

Страница 3: ...t no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the speci cations shall void this warranty IET shall n...

Страница 4: ...ctive 2002 96 EC This product complies with the WEEE Directive 2002 96 EC marking requirements The a xed label indicates that you must not discard this electrical electronic product in domestic househ...

Страница 5: ...ions are taken Make all connections to the instrument before applying power Note the instrument s external markings described under Safety Symbols DO NOT Operate in an Explosive Atmosphere Do not oper...

Страница 6: ...USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when working w...

Страница 7: ...ons for values 100 M 5 3 3 Thermal emf Considerations 5 3 4 Temperature Coe cient Constants 6 3 5 Environmental Conditions 6 3 5 1 Operating Temperature 6 3 5 2 Storage Temperature 6 3 6 Shipping and...

Страница 8: ...lues 190 kW 4 Table 3 1 Connections for values 190 kW 4 Figure 3 2 Connections for values 1 MW and 100 MW 4 Table 3 2 Connections for values 1 MW and 100 MW 4 Figure 3 3 Connections for values 100 MW...

Страница 9: ...s the accuracy by indicating the deviation from nominal for the operating temperature range in 0 5 C increments Because of the low temperature coe cient they require no oil or temperature bath The SRL...

Страница 10: ...5 cm W x 12 7 cm D 3 4 x 4 15 x 5 Weight 0 73 kg 1 6 lb Figure 2 1 OPERATION GUIDE affixed to unit Figure 2 2 Temperature Calibration Chart JOS 19 Jun 2002 SRL 1 B2 9240246 N A 1 000 003 3 N A N A SRL...

Страница 11: ...ppm 4 ppm 2 ppm tot 100 mW 200 mW 500 mW 4 bp s gnd 10 k SRL 10K 2 ppm 4 ppm 1 5 ppm tot 100 mW 200 mW 500 mW 4 bp s gnd 19 k SRL 19K 2 ppm 4 ppm 2 ppm tot 100 mW 200 mW 500 mW 4 bp s gnd 20 k SRL 20...

Страница 12: ...esistance Change 18 C to 28 C 3 ppm Maximum Voltage 0 5 V Maximum Current 0 5 A LO HI LO HI SENSE CURRENT GND HI INPUT HI SENSE LO INPUT LO SENSE GROUND Binding Post Function CURRENT HI Current input...

Страница 13: ...HI GROUND GUARD HI INPUT HI SENSE LO SENSE LO INPUT GROUND GUARD If no GUARD point exists on the measuring instrument it may be connected to GROUND 3 3 Thermal emf Considerations High quality gold pl...

Страница 14: ...ult the Temperature Calibration Chart provided with each unit The calculated resistance value is provided be tween 18 C and 28 C in 0 5 C increments Figure 2 2 shows an example of this table 3 5 2 Sto...

Страница 15: ...e device the following should be observed Calibration Interval General Considerations Required Equipment Calibration Procedure 4 3 1 Calibration Interval The recommended SRL Series calibration interva...

Страница 16: ...precision high stability digi tal multimeter e g Fluke 8508A along with a set of resistance stan dards for ratio mode 4 3 4 Calibration Procedure To calibrate an SRL unit proceed as follows 1 Set up t...

Страница 17: ......

Отзывы: