SRL Series

5

Operation

3.2.3  Connections for values 

³

100 M

W

Values  ≥100  MΩ  have  two  insulated,  low  thermal 
emf binding posts for two-terminal measurements as 
shown in Figure 3-3. The third binding post, labeled 

GROUND

,  is  connected  to  the  case.    The  fourth 

binding  post,  labeled 

GUARD

,  is  connected  to  an 

internal case that contains the resistor.  

Figure 3-3:  Connections for values 

³

100 M

W

Binding Post

Function

HI

Input from source (e.g. ohmmeter)

SENSE LO

Measurement point

GROUND

Shield

GUARD

Interrupts leakage from the internal resistor to 
the case and other components of the unit

Table 3-3:  Connections for values 

³

100 M

W

RESISTANCE STANDARD

MODEL SRL

-

100M

Temperature Coeffi cient: 5 ppm/°C
Maximum Voltage: 1000 V

100 M

W

LO

HI

GROUND

GUARD

HI INPUT

HI SENSE

LO SENSE

LO INPUT

GROUND

GUARD*

*If no 

GUARD

 point exists on the measuring instrument, 

it may be connected to 

GROUND

.

3.3  Thermal emf Considerations

High-quality, gold-plated, tellurium-copper binding 
posts serve to minimize the thermal emf eff ects which 
would  artifi cially  refl ect  a  change  in  dc  resistance 
measurements.    All  other  conductors  within  the 
instrument,  as  well  as  the  solder  used,  contain  no 
metals or junctions that could contribute to thermal 
emf problems.

There nevertheless may be some minute thermal emf 
generated at the test leads where they contact the gold 
banana jacks.  This voltage will also be eliminated if 
a meter with so called “True Ohm” capability is used.  
Otherwise the generated emf may represent itself as a 
false component of the dc resistance measurement.

Always use low emf test leads when working with 
SRL  models.  In  particular,  avoid  brass  or  steel 
conductors.

Содержание SRL Series

Страница 1: ...ASUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC SRL Series Resistance Standard Operation Manual Copyright 2022 IET Labs Inc Visit www ietlabs com for ma...

Страница 2: ...PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC...

Страница 3: ...t no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the speci cations shall void this warranty IET shall n...

Страница 4: ...ctive 2002 96 EC This product complies with the WEEE Directive 2002 96 EC marking requirements The a xed label indicates that you must not discard this electrical electronic product in domestic househ...

Страница 5: ...ions are taken Make all connections to the instrument before applying power Note the instrument s external markings described under Safety Symbols DO NOT Operate in an Explosive Atmosphere Do not oper...

Страница 6: ...USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when working w...

Страница 7: ...ons for values 100 M 5 3 3 Thermal emf Considerations 5 3 4 Temperature Coe cient Constants 6 3 5 Environmental Conditions 6 3 5 1 Operating Temperature 6 3 5 2 Storage Temperature 6 3 6 Shipping and...

Страница 8: ...lues 190 kW 4 Table 3 1 Connections for values 190 kW 4 Figure 3 2 Connections for values 1 MW and 100 MW 4 Table 3 2 Connections for values 1 MW and 100 MW 4 Figure 3 3 Connections for values 100 MW...

Страница 9: ...s the accuracy by indicating the deviation from nominal for the operating temperature range in 0 5 C increments Because of the low temperature coe cient they require no oil or temperature bath The SRL...

Страница 10: ...5 cm W x 12 7 cm D 3 4 x 4 15 x 5 Weight 0 73 kg 1 6 lb Figure 2 1 OPERATION GUIDE affixed to unit Figure 2 2 Temperature Calibration Chart JOS 19 Jun 2002 SRL 1 B2 9240246 N A 1 000 003 3 N A N A SRL...

Страница 11: ...ppm 4 ppm 2 ppm tot 100 mW 200 mW 500 mW 4 bp s gnd 10 k SRL 10K 2 ppm 4 ppm 1 5 ppm tot 100 mW 200 mW 500 mW 4 bp s gnd 19 k SRL 19K 2 ppm 4 ppm 2 ppm tot 100 mW 200 mW 500 mW 4 bp s gnd 20 k SRL 20...

Страница 12: ...esistance Change 18 C to 28 C 3 ppm Maximum Voltage 0 5 V Maximum Current 0 5 A LO HI LO HI SENSE CURRENT GND HI INPUT HI SENSE LO INPUT LO SENSE GROUND Binding Post Function CURRENT HI Current input...

Страница 13: ...HI GROUND GUARD HI INPUT HI SENSE LO SENSE LO INPUT GROUND GUARD If no GUARD point exists on the measuring instrument it may be connected to GROUND 3 3 Thermal emf Considerations High quality gold pl...

Страница 14: ...ult the Temperature Calibration Chart provided with each unit The calculated resistance value is provided be tween 18 C and 28 C in 0 5 C increments Figure 2 2 shows an example of this table 3 5 2 Sto...

Страница 15: ...e device the following should be observed Calibration Interval General Considerations Required Equipment Calibration Procedure 4 3 1 Calibration Interval The recommended SRL Series calibration interva...

Страница 16: ...precision high stability digi tal multimeter e g Fluke 8508A along with a set of resistance stan dards for ratio mode 4 3 4 Calibration Procedure To calibrate an SRL unit proceed as follows 1 Set up t...

Страница 17: ......

Отзывы: