IET Labs Digibridge IEEE-488.2 Interface Board User Manual
Page 12 of 31
IEEE 488.2 / SCPI Digibridge Command Reference
:CONFigure:PARameter
Syntax:
:CONFigure:PARameter {LQ|CD|CR|RQ|LQPPm|CDPPm|CRPPm|RQPPm|
AUTo
|
RX|GB|ZANG|YANG|ZAPPm|YAPPm}
Description:
Sets the measured parameter.
Parameter:
Value
Description
LQ
Inductance
(L/Q)
CD
Capacitance
(C/D)
CR
Capacitance
(C/R)
RQ
Resistance
(R/Q)
LQPPm
Inductance
(L/Q) ppm
CDPPm
Capacitance
(C/D) ppm
CRPPm
Capacitance
(C/R) ppm
RQPPm
Resistance
(R/Q) ppm
AUTo
(default)
Automatic selection of R/Q, C/D, or L/Q
RX
Resistance
(R/X)
GB
Conductance
(G/B)
ZANG
Impedance
(Z/ANG)
YANG
Admittance
(Y/ANG)
ZAPPm
Impedance
(Z/ANG) ppm
YAPPm
Admittance
(Y/ANG) ppm
Equivalent legacy command: M0, M1, M2, M3, M4, M5, M6, M7,
M8
,
K2, K3, K4, K5, K6, K7
Equivalent key:
Note: Parameters in italic are available only on the Digibridge 1693M model.
:CONFigure:CIRCuit
Syntax:
:CONFigure:CIRCuit {PARallel|
SERies
}
Description:
Sets the measurement equivalent circuit.
Parameter:
Value
Description
PARallel
Parallel equivalent circuit
SERies
Series equivalent circuit
Equivalent legacy command: C0,
C1
Equivalent key: