background image

1689-9640 
 

   Page 4 

 

Digibridge IEEE-488.2 Board 

Legacy Digibridge IEEE-488 Commands 

The legacy IEEE-488.1 Digibridge commands are supported by the IEEE-488 interface board for compatibility with 
software and drivers written for the original Digibridge IEEE-488.1 interface boards. 

 

Category 

Description 

Type 

Code 

Display 

Bin # 

2 byte 

D0 

Delta % 

2 byte 

D1 

Value 

2 byte 

D2 

Delta RLC 

2 byte 

D3 

Ratio: nominal value / RLC value 

2 byte 

D4 

Ratio: RLC value / nominal value 

2 byte 

D5 

Bin on ratio: nominal value / RLC value 

2 byte 

D6 

Bin on ratio: RLC value / nominal value 

2 byte 

D7 

Measurement Rate 

Fast 

2 byte 

S0 

Medium 

2 byte 

S1 

Slow 

2 byte 

S2 

Measurement Voltage 

Constant, off 

2 byte 

Y0 

Constant, on 

2 byte 

Y1 

Value (in volts) 

 v 

Floating point 

Vv; 

Function 

Measure 

2 byte 

P0 

Enter 

2 byte 

P1 

Reset 

Turn-On Conditions 

2 byte 

P2 

Diagnostics 

2 byte 

P3 

Measurement Mode 

Triggered 

2 byte 

T0 

Continuous 

2 byte 

T1 

Triggered median value 

2 byte 

T2 

Continuous median value 

2 byte 

T3 

Parameter 

Inductance 

(L/Q) 

2 byte 

M0 

Capacitance  (C/D) 

2 byte 

M1 

Capacitance  (C/R) 

2 byte 

M2 

Resistance 

(R/Q) 

2 byte 

M3 

Inductance 

(L/Q) ppm 

2 byte 

M4 

Capacitance  (C/D) ppm 

2 byte 

M5 

Capacitance  (C/R) ppm 

2 byte 

M6 

Resistance 

(R/Q) ppm 

2 byte 

M7 

Automatic selection of R/Q, C/D, or L/Q 

2 byte 

M8 

Resistance 

(R/X) 

2 byte 

K2 

Conductance  (G/B) 

2 byte 

K3 

Impedance 

(Z/ANG) 

2 byte 

K4 

Admittance 

(Y/ANG) 

2 byte 

K5 

Impedance 

(Z/ANG) ppm 

2 byte 

K6 

Admittance 

(Y/ANG) ppm 

2 byte 

K7 

Equivalent Circuit 

Parallel 

2 byte 

C0 

Series 

2 byte 

C1 

Содержание 1689-9640

Страница 1: ...899 8438 FAX 516 334 5988 www ietlabs com Long Island NY Email info ietlabs com IET LABS INC Copyright 2011 IET Labs Inc Visit www ietlabs com for manual revision updates Digibridge IEEE 488 2 im Marc...

Страница 2: ...PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com Long Island NY Email info ietlabs com IET LABS INC...

Страница 3: ...at no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the speci cations shall void this warranty IET shall...

Страница 4: ...E USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when working...

Страница 5: ...APERture 13 CONFigure MODe 13 CONFigure DISPlay 13 CONFigure DISPlay RATio 14 CONFigure DISPLay BINRatio 14 CONFigure FUNCtion 14 CONFigure RANGe 14 CONFigure DELay 15 CONFigure ITFactor 15 CONFigure...

Страница 6: ...te 23 SYSTem COMMunicate SERial PARity 23 SYSTem COMMunicate SERial STOPbits 23 SYSTem COMMunicate SERial HANDshake 24 SYSTem ERRor NEXT 24 SYSTem VERSion 24 STATus OPERation EVENt 24 STATus OPERation...

Страница 7: ...andshake Talker T5 Talker with normal and talk only modes for systems without a controller switch selectable on the rear panel L4 Listener SR1 Service Request to request service when measurement is co...

Страница 8: ...7 ON ON OFF OFF OFF 8 ON OFF ON ON ON 9 ON OFF ON ON OFF 10 ON OFF ON OFF ON 11 ON OFF ON OFF OFF 12 ON OFF OFF ON ON 13 ON OFF OFF ON OFF 14 ON OFF OFF OFF ON 15 ON OFF OFF OFF OFF 16 OFF ON ON ON ON...

Страница 9: ...Since multiple safeguards are in place to prevent unintentional flash overwriting the use of JP3 is not needed All other headers populated and non populated are intended only for manufacturing test o...

Страница 10: ...easurement Voltage Constant off 2 byte Y0 Constant on 2 byte Y1 Value in volts v Floating point Vv Function Measure 2 byte P0 Enter 2 byte P1 Reset Turn On Conditions 2 byte P2 Diagnostics 2 byte P3 M...

Страница 11: ...2 byte X3 RLC 2 byte X4 RLC Bin 2 byte X5 RLC QDR 2 byte X6 RLC QDR Bin 2 byte X7 RLC QDR Bin on FAIL only 2 byte X8 Same on FAIL Bin on PASS 2 byte X9 Nominal Value Value in ohms henries farads n Fl...

Страница 12: ...e Q4 Quick acquisition shorting 2 byte Q5 Display Digit Blanking Eliminate a digits from RLC b digits from QDR 3 byte Oab Integration Time Multiplying Factor Value i multiplies MEDIUM and FAST integra...

Страница 13: ...ANG Admittance Y ANG Impedance Z ANG ppm Admittance Y ANG ppm M0 M1 M2 M3 M4 M5 M6 M7 M8 K2 K3 K4 K5 K6 K7 CIRCuit PARallel SERies Selects the measured equivalent circuit Parallel Series C0 C1 APERtu...

Страница 14: ...ric VOLTage Sets the measurement signal voltage CONStant OFF ON Constant voltage off Constant voltage on Y0 Y1 LEVel numeric Real value in volts V numeric CURRent numeric Sets constant current Real va...

Страница 15: ...RLC QDR RLC QDR Bin RLC QDR Bin on FAIL only Same on FAIL Bin on PASS X0 X1 X2 X3 X4 X5 X6 X7 X8 X9 DATaformat ASCii BINary Selects GPIB RS 232 output data format Normal ASCII not compacted Compacted...

Страница 16: ...ing GPIB BAUDrate 9600 19200 38400 57600 115200 Sets the RS 232 interface baud rate PARity NONe EVEN ODD Sets the RS 232 interface parity No parity Even parity Odd parity STOPbits 1 2 Sets the RS 232...

Страница 17: ...a measurement like start switch G0 CLS Clear Status Register ESE numeric Sets Standard Event Status Enable ESE Reads Standard Event Status Enable ESR Reads the Standard Event Status Register IDN Read...

Страница 18: ...tance C R ppm RQPPm Resistance R Q ppm AUTo default Automatic selection of R Q C D or L Q RX Resistance R X GB Conductance G B ZANG Impedance Z ANG YANG Admittance Y ANG ZAPPm Impedance Z ANG ppm YAPP...

Страница 19: ...cription Set the measure mode Parameter Value Description TRIGgered Set measurement mode to triggered Continuous Set measurement mode to continuous MTRiggered Set measurement mode to triggered median...

Страница 20: ...panel to display measurement bin on ratio Parameter Value Description NOMValue Display Bin on ratio nominal value RLC value VALNominal Display Bin on ratio RLC value nominal value Equivalent legacy c...

Страница 21: ...lay time Parameter Value Description numeric Delay time real value in ms Equivalent legacy command J numeric Equivalent key CONFigure ITFactor Syntax CONFigure ITFactor numeric Description Set the int...

Страница 22: ...mmand F numeric Equivalent key none CONFigure VOLTage CONStant Syntax CONFigure VOLTage CONStant OFF ON Description Set the measurement signal voltage Parameter Value Description OFF Set constant volt...

Страница 23: ...Disables the DC bias On or 1 Enables the DC bias Equivalent legacy command U0 U1 Equivalent key CONFigure ROUTine Syntax CONFigure ROUTine NORMal REVersal SHORting REVShortint QUICk QSHorting Descript...

Страница 24: ...tion Set display digit blanking Eliminate a numeric digits from RLC and b numeric digits from QDR displays on Digibridge front panel Parameter a numeric Value Description 0 to 5 Number of digits to el...

Страница 25: ...in low limit in Parameter Value Description numeric Real value low limit in Equivalent legacy command L numeric Equivalent key CONFigure BINSummary Syntax CONFigure BINSummary DISable ENABle RESet Des...

Страница 26: ...yntax CONFigure DATaformat ASCii BINary Description Selects the GPIB RS 232 output data transfer format Parameter Value Description ASCii Set output format to normal ASCII text BINary Set output forma...

Страница 27: ...on to turn on conditions Equivalent legacy command P2 Equivalent key SYSTem RESTart Syntax SYSTem RESTart Description Resets the Digibridge and re runs power on diagnostics Equivalent legacy command P...

Страница 28: ...prevent interference with or slowing down of measurements if using another interface such as the RS 232 serial interface for communicating with the Digibridge instrument Parameter Value Description E...

Страница 29: ...nd none Equivalent key none SYSTem COMMunicate SERial PARity Syntax SYSTem COMMunicate SERial PARity NONe EVEN ODD Description Set the RS 232 serial interface parity Parameter Value Description NONe N...

Страница 30: ...andshake Equivalent legacy command none Equivalent key none SYSTem ERRor NEXT Syntax SYSTem ERRor NEXT Description Reads the error message Query only Equivalent legacy command none Equivalent key none...

Страница 31: ...ERation ENABle reads the current value of the Operation Enable Mask Register without clearing the register value Parameter Value Description 0 to 255 Numeric value range Equivalent legacy command none...

Страница 32: ...umeric value range Equivalent legacy command none Equivalent key none STATus PRESet Syntax STATus PRESet Description Enables all required event reporting Equivalent legacy command none Equivalent key...

Страница 33: ...Numeric value range Equivalent legacy command none Equivalent key none ESR Syntax ESR Description Reads out the value of the Standard Event Status Register Executing this command clears the Standard...

Страница 34: ...quivalent legacy command none Equivalent key none RST Syntax RST Description Resets Digibridge settings to Turn On Conditions Equivalent legacy command P2 Equivalent key SRE Syntax SRE numeric SRE Des...

Страница 35: ...ST Syntax TST Description Initiates self test and returns 0 if all tests passed or error code if any test failed Equivalent legacy command none Equivalent key none WAI Syntax WAI Description Waits unt...

Отзывы: