1689-9640
Digibridge IEEE-488.2 Board
Page 5
Range Control
Hold range
2 byte
R0
Hold range 1
2 byte
R1
Hold range 2
2 byte
R2
Hold range 3
2 byte
R3
Hold range 4
2 byte
R4
Autorange
2 byte
R5
Frequency
Value (in kHz)
═
f
Floating point
Ff;
Zero Calibration
Disable
2 byte
Z0
Enable open
2 byte
Z1
Enable short
2 byte
Z2
Output Format
Normal
(not compacted)
2 byte
x0 *
Compacted binary format
2 byte
x1 *
Data Output
None
2 byte
X0
Bin #
2 byte
X1
QDR
2 byte
X2
QDR, Bin #
2 byte
X3
RLC
2 byte
X4
RLC, Bin #
2 byte
X5
RLC, QDR
2 byte
X6
RLC, QDR, Bin #
2 byte
X7
RLC, QDR, Bin # on FAIL only
2 byte
X8
Same on FAIL; Bin # on PASS
2 byte
X9
Nominal Value
Value (in ohms, henries, farads)
═
n
Floating point
Nn;
Limit Entry
Bin #
3 byte
Bbb
High limit (in %)
═
h
Floating point
Hh;
Low limit (in %)
═
l
Floating point
Ll;
Initiation
Start a measurement (like start switch)
2 byte
G0
Manual Start
Enable switch
2 byte
W0
Disable switch
2 byte
W1
Keyboard Lock
Unlock
2 byte
K0
Lock
2 byte
K1
Average
# of measurements
═
a
Floating point
Aa;
Internal Bias
Off
2 byte
U0
On
2 byte
U1
Bin Summary
Disable output
2 byte
E0
Enable output
2 byte
E1
Reset bin summary to 0
2 byte
E2