Service Manual NeuroLight / AlgiScan - Rev B
9
Silicone bumper
Silicone bumper
Product name sticker
3) Unscrew the 6 screws on the front panel using a phillips screwdriver.
4) Disassemble the 2 parts of the handle.
5) Remove the K6 jumpers and disconnect the K2 battery.
b. LEMO connector change (only for the AlgiScan)
1) Unsolder the 2 wires in K1 and then resolder the LEMO connector supplied.
2) Then solder the n°1 wire on the plus, the n°2 wire on the minus.
3) To reassemble the unit follow the previous steps in reverse order.
Note: Both silicone bumpers must be glued with silicone adhesive.