Service Manual NeuroLight / AlgiScan - Rev B
12
The 2 leads have to be soldered to match the following pins configuration:
Pin 1
Pin 6
Pins configuration
Connector pins
Cable Leads
1
White
2
Brown
Solder first the white cable.
Solder all cables counter-clockwise.
Test the proper functioning of the connector using a patient simulator.
Note: If the results are not satisfying please check the soldering.
Apply a drop of glue on the cable collet, then insert the backnut inside the cable collet.
Screw the connector at the max and swipe out the excess of glue.
Apply a drop of silicon glue on the shell of the LEMO connector to keep the blue bend relief in place
(ex.Silcoset 153).
Apply here
Let dry for 24 hours.