Service Manual NeuroLight / AlgiScan - Rev B

11

e. Disassembling mother board (for screen change)

1) Carefully remove the glue from the wireless charging coil

2) Unscrew the 4 screws that hold the motherboard.

3) Disconnect the connection ribbon of the screen and touch pad (x4 and x5)

4) Remove the screen with relatively strong pressure (Caution: wear safety gloves)

5) Remove the seal and replace it

6) Connect the new display to the motherboard

7) Place the screen previously connected to the motherboard on the seal after removing the film 

from the it.

8) Replace the motherboard and put the screws back in place.

f. Wireless charging coil change

1) Same as section 1) disassembling mother board

2) Same as section 2) disassembling mother board

3) Unsolder the coil

4) Solder the new

5) Screw the motherboard back on

6) Place the coil in its slot

7) Apply 3 points of hot glue to fix it

g. Disassembly of the push/pull LEMO connector

Remove the blue cable bend relief then slide it on the cable.

Cut the cable as close as possible from the connector.

To re-do the connector please see the instruction below “Mounting sensor cable push/pull  

connector”.

h. Mounting sensor cable push/pull connector

Connector mounted with its blue cable bend relief

In order to mount the push pull connector, first insert the blue cable bend relief, then mount and sold 

the parts in the following order:

Backnut

Cable collet

Ins

contacts

Shell

  1 

          2   

        4

Remove the sheath between 1cm and 1.4cm (1/2inch).

Cut all wires and shielding except the white and brown wires.

Strip these two wires about 3mm.

Tin all wires.

Содержание Algiscan

Страница 1: ...Service Manual NeuroLight AlgiScan Last Update 2021 04 15 Ref NL ALG SM RevB EN SERVICE MANUAL...

Страница 2: ...Service Manual NeuroLight AlgiScan Rev B 2...

Страница 3: ...controls 5 Maintenance operations specific to the AlgiScan 6 a Hardware part control 6 b Functionals controls 6 II Disassembling process 8 a Disassembling the housing 8 c Battery change 10 d Disassemb...

Страница 4: ...used Check the pupil calibration see sheet final test ENR 13 20 D with barcode Necessary accessories The luxmeter ref LM100 AMPROBE The Pupil Target card The rigid eyecup specific for test b Hardware...

Страница 5: ...on the pupil target card by placing the eyecup on it Check the size that appears on the screen The pupil size should be 3 mm with an accuracy of 0 1mm otherwise continue with the following steps Scan...

Страница 6: ...lses generated by AlgiScan 1 Necessary material AlgiScan to test 1 CAB STIM3 1 oscilloscope probe 1 impedance 4K test card 1 Pupil Target card The rigid eyecup specific for test Rigid eyecup Pupil Tar...

Страница 7: ...n the Pupil Target card with the AlgiScan by placing the rigid eyecup on it Launch a PPI by pressing and holding the central icon for several seconds until all the beeps sound At each audible beep the...

Страница 8: ...st ENR 13 20 D II Disassembling process a Disassembling the housing 1 For the AlgiScan only unscrew the LEMO connector located on the bottom silicone bumper 2 For all devices with a plastic stylus rem...

Страница 9: ...rts of the handle 5 Remove the K6 jumpers and disconnect the K2 battery b LEMO connector change only for the AlgiScan 1 Unsolder the 2 wires in K1 and then resolder the LEMO connector supplied 2 Then...

Страница 10: ...p 3 Reposition the battery holder being careful not to overtighten the screws 4 Reconnect the battery and the jump and reassemble the unit d Disassembling optical room 1 Unscrew the 6 screws on the op...

Страница 11: ...ame as section 2 disassembling mother board 3 Unsolder the coil 4 Solder the new 5 Screw the motherboard back on 6 Place the coil in its slot 7 Apply 3 points of hot glue to fix it g Disassembly of th...

Страница 12: ...lockwise Test the proper functioning of the connector using a patient simulator Note If the results are not satisfying please check the soldering Apply a drop of glue on the cable collet then insert t...

Страница 13: ...Mechanical Drawing Overall set AlgiScan and NeuroLight 8 1 Front housing 2 Back housing 3 2 5x1 6mm countersunk screw 4 Product name sticker 5 Front silicone bumper 6 Rear silicone bumper 7 LEMO conn...

Страница 14: ...Service Manual NeuroLight AlgiScan Rev B 14 Front sub set 1 Front housing 2 Screen seal 3 Touch screen 4 barcode reader 5 2 5mm x 10mm screw...

Страница 15: ...Service Manual NeuroLight AlgiScan Rev B 15...

Страница 16: ...Copyright 2021 IDMED SAS All rights reserved No information or part of this document may be reproduced or transmitted in any form or by any means without the prior written permission of IDMED...

Отзывы: