VAC-5000HD User Guide
43
10 Troubleshooting
In case this device does not work correctly, please check the following items first.
・
Are this device and the connected devices turned on normally?
・
Are cables connected correctly?
・
Are there no loose connections?
・
Are cables that are appropriate to this device being used?
・
Are signal specifications of connected devices matched to each other?
・
Are settings of the sink device correct?
・
Are there any close objects that may cause noise?
If the problem still cannot be solved, check the items in the table below. Please refer to manuals of connected
devices as well, since they may possibly cause the problem.
Problem
Cause/Check item/Solution
Page
●
Video output
Video is not output.
If there are no problems with cable connections, first check [1] and
[2] below.
-
[1]
Verify that the EDID resolution setting of the VAC is set to the
input resolution supported by the display device.
・
The EDID resolution is set to 1080p by default, but some
TVs do not support that resolution.
・
If the EDID resolution is set to 1080i, the video may not be
output to the display device that does not support the
interlaced signals.
・
Some monitors for PCs do not support the resolutions for
TVs. Some LCD TVs do not support PC resolutions (VGA
to QWXGA).
[2]
Verify that signals are output from the input device.
If the “SIGNAL” LED lights, check [3] to [6]; if the LED turns
off, check and [8].
[3]
If signals protected by HDCP are input, does the sink device
support the HDCP? If the sink device does not support HDCP,
copyright-protected signals are not displayed.
Check whether the sink device supports the HDCP.
You can also disable HDCP output to the source device by an
input setting.
[4]
If a long cable is connected for input or output, replace it with
a 5 meter/16.4 feet or shorter cable. A 16 ft. (5 m) or longer
cable can be connected for digital input/output for the VAC,
but some cables fail HDCP authorization or EDID acquisition
depending on the quality of the cable and connected devices.
-
[5]
Verify that unsupported signals are not being input.
[6] Change the [C10] setting. Access this menu by changing
[F99] setting.