Operator Manual
Aspect 100RT
L100-3061 REV A
13
•
RADIATION SHIELDS, SADDLE STYLE (OPTIONAL)
•
24X80 Straight Tabletop Shield
Part Number: A610-096
INSTALLATION INSTRUCTIONS FOR 24” SADDLE STYLE RADIATION SCATTER SHIELD
(0.5mm PbEq)
Note: Individual shields have hook fasteners that are engaged into the loop fasteners of the
saddle. Overlap shields as shown.
Step A: Place the saddle with the loop fastener side “up” on the tabletop. The adhesive dots may be applied
to hold the saddle in place.
A B C D
Step B: Attach two 8 1/2” panels at the end of the tabletop centered as shown.
Step C: Attach two 13” panels so they wrap around the corners.
Step D: Finish by attaching five 11” panels along the side of the tabletop.