Operator Manual        

Aspect 100E 

 

 

L100-3059 REV A 

 

www.imagediagnostics.com

 

                 18 

 
 

MAINTENANCE 

 

”Qualified Technician” 

 

 All maintenance procedures should be done by an experienced technician with demonstrated  

  knowledge and skills (electrical and mechanical) relative to this type of equipment. 

 

 This individual must have access to this manual and the proper tools. 

 

ROUTINE MAINTENANCE

 

• Lubrication is 

not

 required however powdered graphite may be used on the lifting tube. 

 

 

PERIODIC PERFORMANCE CHECKS

 

 

 Semiannually perform functional inspections of carbon fiber tabletop, support frame, 

   and the elevating tube. 
 

• 

Daily inspect all external cables, controls, and the tabletop for wear and damage.   

  Replace damaged cables promptly. 
 

• 

Weekly check battery operation by disconnecting the AC power and running the tabletop 

  up and down. 
  Check daily if table is routinely used without being attached to mains.

 

 

 

Содержание Aspect100E

Страница 1: ...Aspect100E Mobile Imaging Table Operator Manual Made in the USA...

Страница 2: ...T 2 SETUP 3 SAFETY HAZARDS 4 SAFETY HAZARD ALERTS 5 SAFETY INSTRUCTIONS 5 ACCESSORIES 8 MAINTENANCE 18 USER REPLACEMENT ITEMS 19 TROUBLE SHOOTING 19 DISPOSAL OF COMPONENTS 20 WARRANTY 21 Linak documen...

Страница 3: ...REV A www imagediagnostics com 1 PRODUCT DATA Low attenuation carbon fiber tabletop aluminum equivalency is less than 1mm Locking swivel casters Low attenuation table pads Hand operator control is st...

Страница 4: ...ties This equipment should only be installed in an area provided with the proper electrical power The owner is responsible for verifying continued compliance with all applicable regulations and standa...

Страница 5: ...lified technical personnel to repair those parts of equipment which are designated by the manufacturer as repairable For technical assistance call IDI at 978 829 0009 or send fax to 978 829 0027 Be pr...

Страница 6: ...follow these instructions can cause accidents resulting in serious personal injury or damage to equipment CD ROMs containing file copies of all relevant drawings BOMs and documentation are included w...

Страница 7: ...ex Single Use Item This symbol when used in this manual indicates the potential of exposure to harmful x rays Be sure to read and comply with all warnings European Authorized Representative Warning In...

Страница 8: ...50Hz Duty Cycle 10 6 min hr Current Rating Less than 10 Amps Internally Battery Powered Environmental Operating Temperature Range 10 C to 40 C Operating Humidity Range 30 to 75 relative humidity nonc...

Страница 9: ...nt must be installed and used according to the manufacturer s instructions in order to avoid radio frequency interference If this equipment generates or receives interference Verify that the equipment...

Страница 10: ...ositioning Hand controls are standard with this table with foot controls as optional The operator should become familiar with the controls before using The table operates on 110Vac or 230Vac or on int...

Страница 11: ...properly cleaned between patient use as it will inevitably come into contact with contaminants during procedures SAFETY HAZARDS Operators using this equipment should understand the safety issues emer...

Страница 12: ...as not been tested for use with high frequency surgical equipment cardiac defibrillators or cardiac defibrillator monitors WARNING If an antistatic path is required use the equipment on an antistatic...

Страница 13: ...found marked on the equipment means Attention consult accompanying documentation WARNING Always safely position and secure patient on table WARNING Do not exceed patient weight of 450 pounds 204kg CA...

Страница 14: ...operation CAUTION Do not sit on the table top at the head of the table CAUTION Secure wheel locks by depressing locks on all 4 wheels prior to patient usage and when table is not in motion Disengage...

Страница 15: ...IES CRESCENT FACE CUSHION supplied with some models Use this pad for patient comfort when the patient will be positioned face down Place a disposable cover over the pad LATEX Crescent Face Cushion Hea...

Страница 16: ...THE TABLETOP OVER THE TOP SURFACE OF THE TABLETOP PHOTO OF STRAP ARRANGEMENT Patient Restraint Strap Set Standard Part Number C000 0328 Two straps one 3 wide and one 5 wide with hook loop fasteners N...

Страница 17: ...pen latch to install to accessory rail or remove from accessory rail B Armboard pad C Pull trigger to rotate armboard to desired position D Hook loop patient straps Clamp on Rail Accessory Assembly A1...

Страница 18: ...al shields have hook fasteners that are engaged into the loop fasteners of the saddle Overlap shields as shown Step A Place the saddle with the loop fastener side up on the tabletop The adhesive dots...

Страница 19: ...ers that are engaged into the loop fasteners of the saddle Overlap shields as shown Step A Place the saddle with the loop fastener side up on the tabletop The adhesive dots may be applied to hold the...

Страница 20: ...produces ionizing radiation all radiation safety standards and precautions should be applied including but not limited to proper use of X ray protective shielding for patients operators and support p...

Страница 21: ...batteries and achieve maximum battery life If power to table has been absent for an extended period of time battery will need to be recharged The battery is sealed and requires no other maintenance T...

Страница 22: ...desired position TABLETOP MOTIONS UP Depressing and maintaining the UP button on the remote or UP foot switch or toggling the panhandle rocker switch to the UP position causes upward tabletop motion...

Страница 23: ...information about batteries such as battery life production codes and disposal go to the Linak web site for disposal information Select Products User Manuals and click on Actuators Electronics Linak...

Страница 24: ...y access fluids with a water dampened cloth or sponge 5 Clean any accessories tabletop and frame using an approved cleaner 6 Thoroughly rinse tabletop pad with water 7 Gently rub with a soft clean clo...

Страница 25: ...roper tools ROUTINE MAINTENANCE Lubrication is not required however powdered graphite may be used on the lifting tube PERIODIC PERFORMANCE CHECKS Semiannually perform functional inspections of carbon...

Страница 26: ...omply to California Technical Bulletin 117 as well as Underwriters Laboratories 94 UL94 Standard Tabletop Pad Part Number X100 1742 LATEX Not made with natural rubber latex TROUBLE SHOOTING SYMPTOM PO...

Страница 27: ...components typically found in IDI products but varies with model and options COMPONENT ITEM RECYCLING GROUP Actuator Spindle and Motor Housing Cable Metal Steel and Copper Plastic Copper Control Box P...

Страница 28: ...cs com 21 WARRANTY Warranty details for IDI Products can be obtained directly from Image Diagnostics Inc Image Diagnostics Inc 310 Authority Drive Fitchburg MA 01420 USA Or call IDI at 978 829 0009 or...

Отзывы: