Operator Manual
Aspect 100-4
L100-3316 Rev. A
29
Guidance and Manufacturer’s Declaration Emissions, Equipment and Systems that
are NOT Life-Supporting
Table 3
The
Aspect
100-4 is
intended for use in the electromagnetic environment specified below.
The customer or user of the
Aspect
100-4
should ensure that it is used in such an
environment.
Immunity Test
EN/IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Conducted RF
EN/IEC 61000-4-6
Radiated RF
EN/IEC 61000-4-3
AC Mains:
3V, 0.15MHz – 80MHz
6V in ISM band between
0.15MHz and 80MHz
80%AM at 1KHz
[see table 5 of IEC
60601-1-2]
SIP/SOPS:
3V, 0.15MHz – 80MHz
6V in ISM band between
0.15MHz and 80MHz
80%AM at 1kHz
[see table 5 of IEC
60601-1-2]
3V/m
80MHz to 2.7GHz
80%AM at 1kHz
AC Mains:
3V, 0.15MHz – 80MHz
6V in ISM band
between
0.15MHz and 80MHz
80%AM at 1KHz
[see table 5 of IEC
60601-1-2]
SIP/SOPS:
3V, 0.15MHz – 80MHz
6V in ISM band
between
0.15MHz and 80MHz
80%AM at 1kHz
[see table 5 of IEC
60601-1-2]
3V/m
80% AM
Portable and mobile
communications equipment
should be separated from the
Aspect
100-4
by no less than the
distances calculated/listed below:
D=(3.5/3)(Sqrt P)
D=(3.5/3)(Sqrt P)
80 to 800 MHz
D=(7/3)(Sqrt P)
800 MHz to 2.7 GHz
where P is the max power in watts
and D is the recommended
separation distance in meters.
Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey,
should be less than the compliance
levels (V1 and E1).
Interference may occur in the
vicinity of equipment containing a
transmitter.