Operator Manual
Aspect 100-4
L100-3316 Rev. A
30
Recommended Separations from the
Aspect 100-4
are intended for use in the electromagnetic
environment in which radiated disturbances are controlled. The customer or user of the
Aspect
100-4
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF Communications Equipment and the
Aspect 100-4
as
recommended below, according to the maximum output power of the Communications
Equipment.
Table 4
Max Output
Power
(Watts)
Separation (m)
150kHz to 80MHz
D=(3.5/3)(Sqrt P)
Separation (m)
80 to 800MHz
D=(3.5/3)(Sqrt P)
Separation (m)
800MHz to 2.5GHz
D=(7/3)(Sqrt P)
0.01
.1166
.1166
.2333
0.1
.3689
.3689
.7378
1
1.1666
1.1666
2.3333
10
3.6893
3.6893
7.3786
100
11.6666
11.6666
23.3333
17.
WARRANTY
Warranty details for IDI Products can be obtained directly from Image Diagnostics, Inc.
Image Diagnostics, Inc.
310 Authority Drive
Fitchburg, MA 01420 USA
Or call
IDI
at (978) 829-0009.
Recommended Separation Distances between portable and mobile RF
Communication Equipment and the Aspect 100-4. Equipment and Systems that are
NOT Life-Supporting.